A 1588   

                               

            

 Multi range AC current clamps  

 

 

Power Measurement 

A 1588 current clamps have small phase shift (see Figure 7 and Figure 9) over wide frequency 
range. Therefore they are well suited for:

 

  Power and energy measurements (active, reactive, apparent) 

  Power factor measurements 

  Power/Energy efficiency 

Leakage Current Measurement 

Measurement Of Leakage Current By Direct Method 

This method is used when measuring the current through one (embraced) conductor. 

 

Figure 4: 

Leakage current measurement 

– direct method

 

Measurement Of Differential Current And/or Current Difference 

The  differential  method  sums  the  current  passing  through  two  or  more  active  (embraced) 
conductors.  If  no  current  is  leaking  to  earth,,  the  sum  of  the  currents  passing  through  active 
conductors must be exactly zero regardless of the load currents. If a leakage current is flowing 
to earth, it must be equal to the difference in current between the conductors contained in the 
current  clamp  (although,  if  alternative  earth  paths  exist,  this  may  not  be  equal  to  the  current 
passing through the PE conductor). 

Note 

Note  that  the  direction  of  currents  in  the  conductors  must  be  considered  if  the  differential 
method is used. 
 

 

Figure 5: 

Leakage current measurement 

– differential method

 

Determining Problems In Lighting And Grounding Systems. 

Faults  in  electrical  installations  and  equipment  can  cause  additional  continuous  or  short-
lasting leakage currents. Typical faults that can cause increased leakage currents are: 

Summary of Contents for A 1588

Page 1: ...Multi range AC crrent clamps 0 5A 5A 50A A 1588 Instruction manual Version 1 0 Code no 20 752 829...

Page 2: ...el metrel si 2018 METREL Mark on your equipment certifies that this equipment meets the requirements of the EU European Union concerning safety and electromagnetic compatibility regulations No part of...

Page 3: ...MENT 7 POWER MEASUREMENT 8 LEAKAGE CURRENT MEASUREMENT 8 Measurement Of Leakage Current By Direct Method 8 Measurement Of Differential Current And or Current Difference 8 Determining Problems In Light...

Page 4: ...be touched during measurement Symbol on the current clamp indicates the possibility of a hazardous live condition if the operator ignores the required safety measures If the current clamps are used i...

Page 5: ...885 Power Master MI 2892 Clamps have three current ranges 0 5A 5A and 50A which are selected directly by the connected instrument Embedded electronic module is powered directly from the connected inst...

Page 6: ...Master MI 2892 Step 2 Switch on the measuring instrument Set Smart clamps and range on instrument s measuring setup For further information see Instruction manual of measuring instrument Step 3 Clamp...

Page 7: ...rrent primary current Is Measured current current transformer secondary current N2 Number of secondary turns N2 1000 U I Output voltage as function of current U k Im Factor k depends on selected curre...

Page 8: ...e active embraced conductors If no current is leaking to earth the sum of the currents passing through active conductors must be exactly zero regardless of the load currents If a leakage current is fl...

Page 9: ...ve them to dry totally before using it Do not use liquids based on petrol or hydrocarbons Do not spill cleaning liquid over the current clamps To clean jaw cut surfaces use slightly oiled soft cloth S...

Page 10: ...Frequency range 40 Hz 700 Hz 0 50 phase error 700 Hz 2500 Hz 30 phase error Continuity of measurements 100 A continuous 120 A 40 min 20 min intermitted Load impedance 1 k Working voltage 600 Vrms Inf...

Page 11: ...and industrial use Immunity Intended for use in industrial locations but can also be used for domestic purposes Performance criteria B with addition to the tables bellow Table 1 Immunity to radiated...

Reviews: