21
11.3.2021
Metos S29
7. Auto power off setting
Funktionen kan ställas in för att stänga av vågen automatiskt efter en inställd tid när vågen inte används.
1. Press and hold
key until display show F5 OFF during scale is self-checking,
2. Press
key to confirm the selected option.
F5 off
Off
To set auto off function turn off, for scale always on
3 min
Set to turn off three minutes later
5 min
Set to turn off five minutes later
15 min
Set to turn off 15 minutes later
3. Press
key until display show BACK.
4. Press
key to confirm, scale will be start to self-checking and back to normal weighing mode.
6. Calibration
1. Press
key to turn on the scale.
2. 2) Press and hold
key during that start up until display will be shown.
F1 Cal
.
Then you need press calibration switch,which is locating below the battery .
3. Press
to confirm and display will be shown
unlod.
4. Make sure the platter is empty,after stable indicator on
5. Press
, display will be shown current calibration weight value
6. If necessary press
to select the weight value .
7. Place the test weight on the platter.
8. After stable indicator on press
key to confirm.
9. Scale will be restart and come to normal weighing mode.
In case display will show any error message or incorrect measurement, repeat the calibration again.
Kilogram test weights
Model #
(1/3 Weight) Mini-
mum
(2/3 Weight)
(Full capacity) Maximum
S29-3kg
1kg
2kg
3kg
S29-6 kg
2kg
4kg
6 kg
S29-15 kg
5kg
10kg
15 kg
S29-25 kg
10kg
20kg
25 kg