background image

 

 

Troubleshooting 

Octane Basic User Guide 

Page 25 of 25 

 

6

 

TROUBLESHOOTING

 

6.1

 

REMOTE DIAGNOSTICS 

When possible, it may be useful to use a remote access program such as Teamviewer 
to diagnose the system before an on-site visit. Remote access will allow an EDAX 
engineer to examine log files, diagnose system operation and allow service-level 
updates and adjustments to the detector. Using clues from remote diagnostics will 
allow parts to be ordered in advance for an on-site visit if an on-site visit is required.  

 

6.2

 

DETECTOR PROBLEMS 

The detector is connected, and communicates with the software with a network connection. It is 
possible a spontaneous computer glitch can cause a loss of communication with the detector. A typical 
TEAM login may show a 

RED

 indicator for the EDS hardware, as shown below, or display a message 

about DPP service failing to start: 

 

Figure 10 - TEAM Login - RED Hardware Status 

In this case, power off the detector using the switch on the small black power “brick” for the detector 
then reboot the PC. 
 Then power the detector back on when Windows has started. 
 
If the detector shown in the Advanced 
Properties
 is not an Apollo XF, Apollo 
XLT or Octane detector, it could also be 
an indication of loss of communication 
with the detector, as shown in the 
example below. 
 

 

 

Figure 11 - Wrong detector displayed in TEAM 

 

Power off the detector using the switch on the small black power “brick” for the detector. 

 

Reboot the PC. 

 

At the Windows login screen, turn the detector power back on.  

 

Login to Windows. 

 

Check the LED status for the Ethernet connection as per pag

14.

 

 

Wait 1-2 minutes 

 

Launch TEAM or Genesis software 

 

Should be Green 

Incorrect Detector? 

Common fix 
Power PC Off 
Power Detector OFF 
Check cable connections 
Power PC Back ON 
Power the Detector ON 

Summary of Contents for Apollo XF Octane

Page 1: ...Apollo XF XLT Octane Basic User Guide Rev 1 5 June 24 2016...

Page 2: ...uide Page 2 of 25 The Apollo XF Apollo XLT and Octane detectors are manufactured by EDAX a business unit of Ametek Inc Materials Analysis Division 91 McKee Drive Mahwah NJ 07430 USA 201 529 4880 edax...

Page 3: ...Do not expose the detector to extreme temperatures 10 2 6 8 Do not expose the SUTW to plasma 10 2 6 9 Windowless detectors 10 3 Installation and Environment Requirements 11 3 1 Power requirements 11...

Page 4: ...Contents Octane Basic User Guide Page 4 of 25 This page is intentionally left blank...

Page 5: ...r SEM Octane Plus 30 mm2 SUTW for SEM Octane Super 60 mm2 SUTW for SEM Octane Ultra 100 mm2 SUTW for SEM Apollo XF 60 mm2 SUTW for SEM Octane Prime1 10 mm2 SUTW for SEM Apollo XLT 30 mm2 SUTW for TEM...

Page 6: ...Introduction Octane Basic User Guide Page 6 of 25 This page is intentionally blank...

Page 7: ...hot swappable unless otherwise specified The power to the system must be turned off before inserting or removing any of the modules boards or any of the interconnecting cables If this precaution is n...

Page 8: ...were introduced to the EDS market several years ago to improve the transmission of low energy X rays e g B C N O through the window The SUTW is constructed of a 300 nm thick polymer foil with an addi...

Page 9: ...oose particles Avoid venting the sample chamber too quickly Specifically do not exceed 10 cm s gas velocity into the chamber as this can increase the risk of particle induced detector window damage Th...

Page 10: ...indow or may cause the temperature at the window to exceed manufacturer s specifications This may occur during a microscope chamber bake or operation of a heat stage Recommendations If there is a poss...

Page 11: ...5 watts 100 240 Volts 47 63 Hz 0 5 to 1 25 Amps Table 2 Power requirements 3 2 SPACE AND WEIGHT SPECIFICATIONS Workstation dimensions 17 5 H x 7 75 W x 19 5 in 44 x 20 x 49 5 cm Workstation weight 33...

Page 12: ...Installation Requirements Octane Basic User Guide Page 12 of 25 This page is intentionally blank...

Page 13: ...sis software By default there is an Administrator login set up for TEAM Figure 1 Detector Power Supply 4 2 DETECTOR COOLING By default detector cooling is Off until it is started by the user When the...

Page 14: ...N Ramping GREEN At Operating Temp Blinks RED Warming Table 4 Status LEDs definitions When the detector is first powered ON both Status lights on the detector back panel will light RED There is a Yello...

Page 15: ...EAM Status Indicator Green Red OK to Vent DU is warm OK to vent microscope DU is cold Not OK to vent microscope for Window less DU Ready DU is cold OK to collect spectra DU is not ready to collect spe...

Page 16: ...e Some or all may simply be software buttons that the user must acknowledge that it is ok to cool the detector Toggle so each Safety Interlock is Green Figure 4 Detector control and detector status fo...

Page 17: ...n This high count condition is normally set by the factory or installation service engineer The purpose is to protect the detector from high energy backscattered electrons in a TEM which can be harmfu...

Page 18: ...ut for optimal performance and stable peak positions it may be best to wait about 60 minutes It is okay to leave the detector cold when the microscope chamber is left under vacuum There is a user pref...

Page 19: ...5 1 TYPICAL WORKSTATION CONFIGURATION Figure 6 PC Workstation Connections Power Supply Air Flow Hard drive USB Restore SG 3 Board Fan Air Flow Internal Sync cable Sync Panel Network Board Card bracke...

Page 20: ...Ie 2735 171 30939 IP 192 168 0 101 SYNC PCIe 5335 007 28300 SG 2 PCI 5335 007 20000 Mains 24V Power Supply 2735 171 31074 Microscope Octane IP 192 168 0 100 NIC IP Microscope PC IP Ethernet cable 2435...

Page 21: ...l 5335 007 31700 SG 3 PCIe 5335 007 31400 Mains 24V Power Supply 2735 171 31074 Microscope Octane IP 192 168 0 100 NIC IP Microscope PC IP Ethernet cable 2435 072 44045R Sync cable 4035 008 10900 Ferr...

Page 22: ...PCIe 2735 171 30939 IP 192 168 0 101 SYNC Panel 5335 007 31700 SG 3 PCIe 5335 007 31400 Mains 24V Power Supply 2735 171 31074 Microscope Octane IP 192 168 0 100 NIC IP Microscope PC IP Ethernet cable...

Page 23: ...Basic System Cabling Octane Basic User Guide Page 23 of 25 This page is intentionally blank...

Page 24: ...Basic System Cabling Octane Basic User Guide Page 24 of 25 This page is intentionally blank...

Page 25: ...dware as shown below or display a message about DPP service failing to start Figure 10 TEAM Login RED Hardware Status In this case power off the detector using the switch on the small black power bric...

Reviews: