Resetting the Default Sampling Routine 

To restore  all 32 ports to active sampling status, press in turn, “Local”; “Program”; “Enable All 
Ports”; and “Enter” keys.  
To completely reset the manifold controller to factory default settings and clear the memory of all 
programming, turn the power switch on the rear of the controller to the off position, and, while 
holding down the “Program” key, turn it back on.  

Setting Alarm Limits 

The alarm limits for the various particle size channels in the Particle Counter may be copied into the 
Manifold Controller memory so that each sample point may have unique count alarm limits. Figure 
10 and the following example demonstrates the steps to set alarm limits for the system, using the 
sampling routine previously described under “Programming a Detailed Sampling Routine.”  
 
At the 

Manifold Controller

1.  Press “Local” key, verify LED is lit. Two digit 

display will briefly blink the first scan step (00) 
and then display the associated  port number (01) 

2.  Press “Program” key, verify LED is lit. 

 
 

 

At the 

Particle Counter

: Verify the counter is 

displaying the particle size range in the 3 digit LED 
and is not running. 

3. Press 

 or 

 

keys to display the desired 

size range. 

4.  Press “Program” to display alarm limits for 

the selected size range. 

5.  Use the four arrow keys to set the digits to 

the desired alarm limit. 

6.  Press “Enter” to save settings. 
7.  Repeat steps 3 – 6 for each desired size range 

 

 
At the 

Manifold Controller

8.  Press “Copy” key to copy alarm limits from the 

Particle Counter to the Manifold Controller for that 
scan step and port number 

9.  Press “Next” key to advance to the next scan step 

(01) and port number (07) in the sample routine. 
Note: The “Back” key may be pressed to return to 
a previous scan step. 

10. Repeat steps 3 – 9 for the remaining scan steps and 

press “Enter” when all alarm settings have been copied.  

 
If programming of alarm limits in all 32 ports is desired, each port may be selected in turn by using 
the 

↑↑↑↑

 or 

↓↓↓↓

 keys on the manifold controller instead of the “Next” and “Back” keys.  To view the 

Alarm settings for the sampling routine after copying is complete, select the desired scan step using 

 

 

 

 

Figure 10. Copying Alarm Limits 

Summary of Contents for A2432

Page 1: ...Model A2432 Manifold System Operators Manual...

Page 2: ...equire an export license Contact the factory for more information Manual Part Number 701127 Rev Description ECO Date A Original Issue 2 95 B Revision Update 10 97 C Revision Update 6 00 D Firmware cha...

Page 3: ...n Problems 21 Manifold Cleaning and Maintenance 22 Reference Information 23 Manifold Sample Flow 23 Resetting Manifold Controller 22 Scan Table Downloading format 22 Changing Home Test Frequency 22 Sp...

Page 4: ...f electric shock do not expose to rain or moisture WARNING Failure to use this instrument in a manner intended by Pacific Scientific Instruments may circumvent the protection provided by this instrume...

Page 5: ...on you must first contact your local service representative or Customer Service for return authorization instructions After you have received a Return Authorization number follow the shipping instruct...

Page 6: ...lifornia Avenue Grants Pass OR 97526 TEL 541 479 1248 800 866 7889 FAX 541 479 3057 Eastern 127 G Gaither Drive Mt Laurel NJ 08010 TEL 856 235 8224 FAX 856 235 7370 Southern TEL 888 894 1655 FAX 817 6...

Page 7: ...the sampling process automatic Particle Counter Four different particle counters are available for use as part of the manifold system models A2100 A2200 A2400 and A2408 All four have two four five or...

Page 8: ...ning to minimize particle adhesion to walls 3 8 I D for connecting isokinetic probes to manifold larger quantity typical Part number 960380 1 4 I D Hytrel for connecting manifold to particle counter s...

Page 9: ...min 1 0 ft3 min a transit tube of up to 30m may be used For particles in the range of 2 to 10 m the transit tube should be no longer than 3m Under these conditions losses of small particles by diffusi...

Page 10: ...le size distribution at the particle counter because the larger particles are heavier and may drop out in bends or loops Each bend radius should be no less than four inches and care should be taken to...

Page 11: ...nter of shroud 3 Attach a return air tube 1 4 I D between fitting on rear panel of particle counter and manifold 4 Attach large hose between vacuum pump and manifold assembly exhaust outlet leaving un...

Page 12: ...tor and manifold controller rear panel Particle Counter connector 2 Connect the manifold cable between the manifold assembly and controller rear panel connector marked Manifold 3 Plug the counter and...

Page 13: ...e vacuum pump to line power that matches the requirements of the pump in accordance with local electrical codes A wiring diagram is provided in the vacuum pump Operation Manual 5 Check rear panel iden...

Page 14: ...used in a given sampling operation programmed in the particle counter Refer to Table 2 to interpret the following example 1 A sampling operation is set up with 5 sampling points The sampling points a...

Page 15: ...teps programmed 00 31 Particle Counter Cycles set to 1 for one sample at each scan step Locations set to 0 for continuous sampling Sample time set to 1 minute which draws a 1ft3 sample Hold time set t...

Page 16: ...red to specific situation requirements use the following procedure The example given uses the earlier example wherein only 5 sampling points are used and they are connected to manifold ports 01 07 13...

Page 17: ...escribe programming the particle counter for this sampling routine 1 At the Particle Counter Figure 9 press until Cycles LED is Lit 2 Press Program and verify the 3 digit LED displays CYC and the 7 di...

Page 18: ...2 Press Program key verify LED is lit At the Particle Counter Verify the counter is displaying the particle size range in the 3 digit LED and is not running 3 Press or keys to display the desired size...

Page 19: ...The Manifold will begin sampling from the first programmed port and proceed through the sampling routine until it is complete Using the Alarm Module Option Each alarm module provides a remote alarm c...

Page 20: ...d controller 2 Connect the alarm device between the applicable voltage and the desired alarm number output 1 to 16 or 17 to 32 if so connected 3 Connect AC power cable from alarm module to facility po...

Page 21: ...connected Symptom Unable to achieve 1 0 cfm sample flow at the particle counter with rear panel flow adjustment valve Probable Cause Reversed Sample and Return Air tubing connections at the manifold...

Page 22: ...ed to inlet and outlet fittings vacuum hoses may be left attached 2 Remove 12 hex head cap screws holding front panel to manifold body 3 Gently pull front panel straight out and away from manifold 4 U...

Page 23: ...le counter only draws 1 cfm through the sample line and selected port via the sample pickup shoe allowing the excess sample to flow around the pickup shoe This positive pressure and flow around the pi...

Page 24: ...e following string to the manifold controller scan_XX_XX_XXCR where scan command must be lower case _ space 20 hex XX port number CR carriage return 0D hex Changing Home Test Frequency A home test is...

Page 25: ...7 lbs 3 1 kg Program Cycle Up to 100 steps Particle Counter The counter you have selected is identical to the standard counter except that a return air fitting is installed on back panel and a differe...

Page 26: ...lity and Directive 73 23 EEC for Low Voltage Compliance was demonstrated to the following specifications as listed in the official Journal of the European Communities EN 50081 1 1992 Emissions EN 5501...

Page 27: ...t of force majeure iii misuse fault or negligence of or by Buyer iv use of the Products in a manner of which they were not designed v causes external to the Products such as but not limited to power f...

Reviews: