1. Connect a reference flow meter to the inlet fitting on the top of the unit.
2. Access the Set Flow Screen (Figure 21) by pressing Menu then select Set Flow. The
pump will start automatically when you enter the Set Flow Screen and stop when
you leave the screen.
3.
Press ◄ or ► to obtain a reading of 0.1 CFM (2.83 LPM) +/- 5% on the reference
flow meter.
4. Press Enter to save the setting or ESC to cancel the change.
SET FLOW
< > TO ADJUST
ENTER TO SAVE
███████
Figure 21
– Set Flow Screen
5.10. Set Sizes Screen
The GT-526S has six standard factory calibrated particle sizes. These standard sizes
will support most applications and provide the best size accuracy (+/- 10%). This unit
also supports custom sizes. These sizes are configured using the Set Sizes Screen
(Figure 22). Custom sizes are interpolated using the standard size calibration curve.
Therefore, the size accuracy for custom sizes is somewhat reduced (+/- 15%).
SET SIZES
0.3µ 2.0µ
0.5µ 5.0µ
1.0µ 10.0µ
Figure 22
– Set Sizes Screen
The unit sorts sizes from small to large after each size change. Duplicate sizes are not
allowed. Any attempt to set two or more sizes to the same value will result in a
“DUPLICATE SIZES!” warning message.
5.11. Set Clock Screen
To set the date and time select SET CLOCK from the menu. Figure 23 shows the Set
Clock Screen and the following table describes the date / time formats.
SET CLOCK
DATE: 18 AUG’06
TIME: 11:25:36
Figure 23
– Set Clock Screen
Date / Time Formats
Date
dd mmm’yy
dd=day, mmm=month, yy=year
Time
hh:mm:ss
Hh=hours, mm=minutes, ss=seconds