2.3. Test 

Run 

The battery pack inside the GT-321 should be charged for 15 hours prior to use (see 
section 5.3 – Battery Pack).  Once the battery pack is fully charged, 

remove the rubber

 

cap

 from the GT-321 inlet nozzle. The inlet nozzle protrudes from a hole in the top of 

the GT-321 case.  

Next, turn on the power switch that is located on the left side. The LCD display on the 
GT-321 will indicate 

0.3u.  

If there is no display or a dim display the LCD contrast may 

require adjustment (see section 5.6).   

Press the 

<SELECT>

 key several times and note that the display sequences through 

the particle size selections of 

0.3u

0.5u

1.0u

2.0u

, and 

5.0u

. Select the 0.3u size, 

then hold the GT-321 with the inlet nozzle pointing upward.  

Finally, press the 

<START>

 key. You should hear the internal vacuum pump start 

running. After 9 seconds (subsequent readings every 6 seconds) a number will appear 
on the display representing the number of particles per cubic foot larger than 0.3µm.  

If the sample was taken in a relatively clean area the count may be as low as 200,000 
whereas in a dusty area it could be as high as 3,000,000. The GT-321 will continue to 
take 9 more 6-second samples. The results from the 10 samples are then averaged and 
displayed.  

This finishes the test run.  Always remember to turn the power off when not in use to 
conserve battery power.  

The GT-321 is now checked out and ready for use, however, you should read the rest of 
this manual to better understand how to use and maintain this instrument.  

 

 

3. GT-321 

Overview 

3.1. Description 

The GT-321 is designed to be a small, easy to use and completely portable hand-held 
particle counter that can provide fast and accurate measurement of particulate 
contamination in particles per cubic foot. The GT-321 has 5 selectable size ranges of 
0.3, 0.5, 1, 2, and 5 micron.  

The GT-321 contains a laser diode based sensor, NiCd battery pack, vacuum pump, 
microprocessor electronics, LCD display and user input keys all in one small package. 
The accuracy of the GT-321 is certified according to ASTM and JIS standards and 
comparable to larger more expensive bench top particle counters. The fast 6-second 
response of the GT-321 makes it especially useful for troubleshooting contamination 
problems and tracing contamination leaks to the source.  

The laser diode based sensor inside the GT-321 utilizes a specially designed elliptical 
mirror for high light collection efficiency and improved signal to noise. Precision optics 

GT-321-9800 Rev D.doc 

GT-321 Operation Manual 

Page 9 of 18 

AC Tool Supply

877-207-1244

www.aikencolon.com

Summary of Contents for GT-321

Page 1: ...GT 321 HAND HELD PARTICLE COUNTER OPERATION MANUAL GT 321 9800 Rev D doc AC Tool Supply 877 207 1244 www aikencolon com...

Page 2: ...e Instruments Inc Technical Support Should you require support please consult your printed documentation to resolve your problem If you are still experiencing difficulty you may contact a Technical Se...

Page 3: ...nce of procedures other than those specified herein may result in hazardous radiation exposure WARNING This product when properly installed and operated is considered a Class I laser product Class I p...

Page 4: ...ll be warranted to the extent and in the manner warranted by the manufacturer of that product Any product found to be defective during the warranty period will at the expense of Met One Instruments In...

Page 5: ...eration 11 4 1 Iso kinetic Sampling 11 4 2 Taking a Sample 11 5 Maintenance 12 5 1 Service Schedule 12 5 2 Service Schedule Table 12 5 3 Battery Pack 13 5 4 Zero Test 13 5 5 Flow Rate Test 14 5 6 Disp...

Page 6: ...t correctly followed could result in damage to the instrument Do not proceed beyond a caution indicator without understanding the hazard A WARNING indicates a hazard to you and calls attention to a pr...

Page 7: ...ewdriver 4 Iso kinetic Sample Probe with a short piece of tubing 5 Zero Particulate Filter Met One Part Number G3111 6 Universal AC to DC converter module with IEC AC power cord 7 RS 232 Serial Cable...

Page 8: ...batteries Iso kinetic Filter Reduces count errors due to mismatched sample flow velocity Display Adjust Adjusts the LCD display contrast Section 5 6 Flow Adjust Adjusts the flow rate of the unit Secti...

Page 9: ...m the 10 samples are then averaged and displayed This finishes the test run Always remember to turn the power off when not in use to conserve battery power The GT 321 is now checked out and ready for...

Page 10: ...trol Ventilation Systems Filter Testing Laboratory and Manufacturing Environments Concerned citizen groups that want to do their own air quality studies Hospitals and nursing homes Efficiency testing...

Page 11: ...be aware that not all contamination problems are continuous some are the result of a short term event or burst of particles Locating the source will require taking a number of samples in the same spo...

Page 12: ...to the inside of the GT 321 and a factory authorized person must do this Contact Met One Instruments for service information Calibrating particle sensors like the one in the GT 321 requires specialize...

Page 13: ...used powered by its battery pack on a daily basis connect the charger after each day of usage It will not damage the battery pack to leave it connected to the charger for long periods When the GT 321...

Page 14: ...m ID tubing Tygon tubing is a good choice Turn on the GT 321 and note the flow meter reading The flow rate should be 0 10cfm 5 The flow rate can be adjusted by a trim pot located in one of the two acc...

Page 15: ...onnector or defective component inside Send to service center Sample result remains at zero after sampling 1 Pump stopped 2 Laser diode bad 1 Send to service center 2 Send to service center Sample res...

Page 16: ...ter Enable Automatic Records Output a record after each count cycle L Send last record Output the data from the last count cycle Display Commands Output the help menu Example Record CSV Comma Separate...

Page 17: ...Flow rate 0 10 cfm 2 831 lpm Display 16 character LCD Power 6 V NiCd battery pack Charger AC to DC module 9 V 350 mA Operating Temperature 0 to 50 C Storage Temperature 20 C to 60 C Size L 15 9 cm W...

Page 18: ...bers GT 321 GT 321 1 GT 331 GT 521 Product Options All Are in compliance with the following documents EMC Emissions CISPR 11 1990 EN 61326 1 Immunity EN50082 1 EN 61326 1 Tom Pottberg President Januar...

Reviews: