Met One Instruments AEROCET-380 Operation Manual Download Page 22

Page 21 

AEROCET-380 Operation Manual Rev A 

5.5  About Screen 

Figure 5-12 shows the ABOUT screen. The ABOUT 

screen shows the manufacturer’s serial 

number, the firmware version, Met One Instruments, Inc. service contact details, date of last 
calibration, and instrument run time.   

 

Figure 5-12 - About Screen 

 

Charging the Battery 

Caution:  

The provided battery charger is designed to work safely with this device.  Do 
not attempt to connect any other charger or adapter to this device.  Doing so 
may result in equipment damage.  

 

To charge the battery, connect the battery charger to an AC power outlet and the DC plug to 
the socket on the back of the AEROCET-380.  The battery charger is universal and will work 
with power line voltages of 100 to 240 volts, 50 to 60 Hz.  The battery charger LED will be 
Red while charging and the LED will turn Green when the battery is fully charged.  

 

When fully charged the battery inside the AEROCET-380 will power the unit for about 8 
hours of continuous sampling when the backlight is set to timeout.  For continuous operation, 
operate the unit with the battery charger attached.  Charge the battery before storing the 
AEROCET-380.  Storing a discharged battery will degrade its performance. 

 

 

 

About

Aerocet 380

A12345

83604 R1.0.0

[email protected]

541-471-7111

Calibration Date:

Run Time:

10/14/2021
00d 01h 32m

Summary of Contents for AEROCET-380

Page 1: ...0 Particulate Monitor Operation Manual Revision A Met One Instruments Inc Corporate Sales Service 1600 NW Washington Blvd Grants Pass OR 97526 Tel 541 471 7111 Fax 541 471 7116 www metone com service...

Page 2: ...Page 1 AEROCET 380 Operation Manual Rev A...

Page 3: ...tive during normal business hours 7 00 a m to 4 00 p m Pacific Time Monday through Friday Voice 541 471 7111 Fax 541 471 7116 E Mail service metone com Mail Technical Services Department Met One Instr...

Page 4: ...10 4 Operation 11 4 1 About the Measurement 11 4 2 Power Up 11 4 3 Operate Screen 11 4 3 1 Sampling 11 4 3 2 Sample Status 12 4 3 3 Sample History 12 5 Main Menu 13 5 1 Edit Main Menu Items 13 5 2 Sam...

Page 5: ...s 26 9 2 Output 27 9 2 1 Comma Separated Value CSV 27 9 2 2 Status 27 9 3 MODBUS Communication 27 9 3 1 MODBUS 4x Registers 27 9 3 2 MODBUS 3x Registers 28 10 Maintenance 29 10 1 Suggested Periodic Ma...

Page 6: ...peration testing maintaining and troubleshooting your AEROCET 380 unit Should you still require support after consulting your printed documentation we encourage you to contact one of our expert Techni...

Page 7: ...d Conformity The AEROCET 380 is considered a Class I laser product Class I products are not considered to be hazardous This system contains a diode laser operating at 5 mW power and 670 nm wavelength...

Page 8: ...mW 670nm Visible red Display 3 5 color LCD Flow Rate 2 0 liters minute 0 1 LPM Actual volumetric flow Pump Type Brushless Diaphragm 20 000 hour Power Input 14 8V Li Battery Output 16 8VDC 1 8A Power...

Page 9: ...e commercial carrier immediately You should follow any special unpacking instructions provided by the carrier as you then carefully remove all items from the containers and inspect each component It i...

Page 10: ...ort For USB serial communication Pump Filter A 5 micron pump filter keeps any large particles from entering the sample pump Purge Filter A 0 01 micron purge filter cleans the recirculated purge air wh...

Page 11: ...un a purge cycle then a zero cycle then begin sampling 4 Observe the PM level on the display 5 The unit is ready for use 3 User Interface The AEROCET 380 user interface is composed of a rotary dial 2...

Page 12: ...en the unit is started it first enters a Purge Zero cycle where the purge pump is turned on to purge the dirty particles from the engine then it does a Zero calibration The unit will then perform a sp...

Page 13: ...check that the unit can read high and zero concentrations Sampling The AEROCET 380 is currently counting and taking real time samples Stopping Waiting for the sample to finish before coming to a compl...

Page 14: ...on date and run time Figure 5 1 Main Menu Screen 5 1 Edit Main Menu Items To change settings press MENU to display the Main Menu rotate the dial to navigate to the desired item and press the dial to d...

Page 15: ...ime the unit will sample for The sample time is user selectable from 1 minute 2 minutes 5 minutes 10 minutes 15 minutes 30 minutes or 60 minutes 5 2 3 Units This setup feature allows the user to selec...

Page 16: ...in its memory For instructions on viewing stored data see Section 4 3 3 Because this memory is circular once all records are full any new samples taken will overwrite the oldest stored sample data The...

Page 17: ...the sampled particles The individual user k factor can be set from 0 1 to 20 0 The AEROCET 380 calibration is performed using ideal polystyrene latex PSL spheres which provide a powerful tool for ass...

Page 18: ...ength of appropriate flexible tubing The AEROCET 380 flow rate should be maintained to within 0 1 LPM 1 9 to 2 1 LPM for proper cut point performance of inlet cyclones Figure 5 6 Flow Calibration Scre...

Page 19: ...nitial Setup Screen 5 4 1 Serial The Serial setting controls various communication behavior of the AEROCET 380 The following table lists the Serial settings and describes their meanings Selection Desc...

Page 20: ...return to the Initial Setup screen Figure 5 8 Clock Screen 5 4 4 Locations The Locations screen allows the user to set up to 10 predefined alpha numeric locations Each location can use up to 7 charac...

Page 21: ...e each process step An initial warning explains that this is only intended to be used to recover a lost IP Address Select CONTINUE A green bar will show the progress of the scan Figure 5 10 WiFi Disco...

Page 22: ...r to an AC power outlet and the DC plug to the socket on the back of the AEROCET 380 The battery charger is universal and will work with power line voltages of 100 to 240 volts 50 to 60 Hz The battery...

Page 23: ...powering on the AEROCET 380 the WiFi interface will generate a local network that allows the user to connect to it and configure the connection 2 On a WiFi enabled computer or smartphone search for an...

Page 24: ...work it will show the internal IP address it was assigned from the network Click the IP address to automatically navigate to the new IP address The dock will now automatically shut off its hosted netw...

Page 25: ...k 9 Navigate to the uC Console tab and set the baud rate from the drop down tab This must match the WiFi Baud of the measurement device Default WiFi Baud is 115200 Figure 7 5 Microcontroller Console 1...

Page 26: ...ware interfaces The settings baud rate parity and stop bits must match the computer setting for proper communication regardless of the hardware interface type 9 1 1 Computer Mode Computer Mode is inte...

Page 27: ...scriptors DT Date Time YYYY MM DD HH MM SS ID Location ID MA Modbus Address 1 247 MP Modbus Port MP 0 USB MP 1 RS 485 MP 2 WiFi OI Output Interval OI 0 Off OI 1 On PR Print File QH Report data record...

Page 28: ...test failure 1 2 Span Test failure 2 4 IOP Current alarm 3 8 Pressure Sensor failure 4 16 Flow Failure more than 5 for 5 minutes 5 32 Not Used 6 64 Internal Link Down 7 128 Low Battery 9 3 MODBUS Com...

Page 29: ...MB_Inst_Time 1000 dword 2 Current Timestamp Unix MB_Inst_Stat 1002 dword 2 Status Word MB_Curr_Location 1004 string 4 Location String MB_Inst_Conc 1008 float 2 Instantaneous Concentration MB_Inst_Flo...

Page 30: ...ation and optical system cleaning Every 24 months 10 2 Flow Audit and Calibration This procedure is discussed in Section 5 3 4 10 3 PUMP and PURGE Filter Changes The AEROCET 380 PUMP FILTER and PURGE...

Page 31: ...rimarily consists of optical system cleaning laser detector checks and recalibration As found calibration checks can also be requested Contact the Met One technical service department to schedule AERO...

Page 32: ...le result is lower than normal Atypical sample results may be real Flow rate is not 2 LPM Optics may be contaminated Audit and calibrate flow Contact service center Sample result is higher than normal...

Page 33: ...Holder Black Aluminum 80588 Pump Filter Holder Black Aluminum 80588 1 Purge Air Filter 0 01 micron 580302 Pump Filter 5 micron 580292 Included Accessories Li Ion Smart Charger with power cord 81818 1...