9Correct the gain
A higher gain causes a greater slope of the measuring path-output signal characteristic curve
(or rotation of the characteristic curve counterclockwise).
5.4 Setting of transducer type
Before starting the adjustment, the measuring amplifier must be set whether an inductive
displacement transducer with symmetrical or with asymmetrical coil mount
is to be used. In the
standard configuration, the amplifier
is
set for symmetrical displacement transducers. If an
asymmetrical displacement transducer is used, the JP2 jumper must be plugged into the
righthand position.
Figure 7
Selection of Transducer Type (Symmetric/Unbalanced)