10.0 Troubleshooting

Answers to common operational questions

Mesa is ready to help you with any operational issue you may encounter with your De-
fender.  But we may be able to save you some time by providing a short checklist of the 
questions most commonly asked of our customer service and technical specialists.

Why won’t my Defender turn on?

If the Defender will not turn on, verify that the battery has been charged. When connect-

ed to the AC power adapter/charger and power is present a small green indicator light 

should be visible through the front viewing window

My Defender won’t respond to push-button commands.

If the Defender fails to respond to push-button commands, you can perform a hard reset 
of the Defender. This can be done by inserting a paper clip into the reset opening in the 
back of the unit.

I’m not sure I have my Defender connected properly.

 

Verify that the flow source is connected to the pressure port of your Defender for pres

-

sure sources or to the suction port for verifying suction pumps. The unused port should 
be at atmospheric pressure with any cap or plug removed. If you are calibrating a gas 
that requires an exhaust line to vent the measurement gas, ensure that the tubing is of 

sufficient diameter not to create a pressure drop greater then 5 inches of water.

How do I protect against leaks?

Ensure that hose and tube fittings are tight and leak free. The tubing connecting your 
flow source (pump, mass flow controller, needle valve, sonic nozzle or restrictor) to the 

meter should be kept as short as possible. 

What do I do when my leak test fails?

 

 

First check to make sure that the leak test cap is on correctly and it is not leaking 
through the leak test cap itself. If the leak test cap is correct perform leak test both at 
the pressure and suction side. If it fails, contact Mesa Technical Support.

What’s the best way to connect to the filter medium?

When calibrating sampling pumps best results are obtained with the filter medium con

-

nected to the pump and the Defender connected to the inlet side of the filter medium 

with a short piece of tubing. 

Summary of Contents for Defender 530

Page 1: ...Defender 530 User Manual...

Page 2: ...Menu Navigation 2 5 Set Up 2 6 Measurements 2 7 Single 2 8 Burst 2 9 Continuous 2 10 Data Port 3 0 DryCal Pro Software 3 1 Defender Firmware Upgrades 3 2 Annual Maintenance Cal 4 0 Shipping 5 0 Stora...

Page 3: ...uction Operation Data Port Annual Maintenance and Calibration Shipping Storage Defender 530 Specifications Factory Default Settings Limited Warranty Troubleshooting 2 0 6 0 3 0 7 0 4 0 8 0 5 0 9 0 10...

Page 4: ...ded to operate your Defender If at any time you have questions regarding its operation please contact Mesa Labs through our Web site drycal mesalabs com or call 973 492 8400 to speak with a member of...

Page 5: ...3 Data Port Suction Fitting Measuring Cell Pressure Fitting Charging Jack Calibration Label Reset Button Display Anti tamper Label 1 0 3...

Page 6: ...outlet Initial charging should take about eight 8 hours After the initial charge You may continue to charge your Defender indefinitely simply by leaving it connected to the AC power adapter Be sure t...

Page 7: ...lated sealed lead acid battery Please note that opening the Defender may damage connections so this procedure should only be used for battery dis posal Procedure Remove five Phillips head screws on th...

Page 8: ...product name model number and flow range Press the On Off button for 3 seconds to turn your Defender off 2 3 Connections Attaching your Defender to a device Connect device to be calibrated to the app...

Page 9: ...ighlighted for easy identification 2 5 Menu Navigation Moving through operational menus Use the directional arrows and on the control panel to find your way through the menu When your desired command...

Page 10: ...ith DryCal Technology MEASURE SETUP 2 6 Set up Customizing the Defender to your needs You can customize your Defender in the SETUP menu Highlight SETUP in the introduc tion screen to enter the Setup M...

Page 11: ...Hg whereas standardizing temperature is an user settable value set at Std To in the Units sub menu Choose the number of measurements in the average from 1 to 100 If you wish to incorporate a time dela...

Page 12: ...temperature by setting Std To to a value from 0 to 50 deg C or 32 to122 deg F Std To effects the flow rate measurement only when the reading Type in the Reading sub menu is set to standardized Std Ti...

Page 13: ...KC to operate with SKC CalCheck Interface Magnification Controls the amount of data on the display Select Zoom to view only flow measure ments in larger font or select Detail to simultaneously view fl...

Page 14: ...e to misuse or accidental damage Please note that a leak test is not a substitute for a comprehensive examination of the unit s overall per formance and it does not ensure that your Defender is operat...

Page 15: ...t of Range warning appears Immediately lower or disconnect the flow When the flow is within the proper range select RESET to clear your Defender s last measurement About This screen provides basic inf...

Page 16: ...for 1 second to turn on your Defender When first turned on your Defender displays an opening screen showing the product name model number and flow range Press the On Off button for 3 seconds to turn...

Page 17: ...urement This setting functions in the same manner as SINGLE but measurements continue automatically until the preset number of measurements has been made Operation then ceases and the last reading and...

Page 18: ...to 100 by accessing the SETUP menu In Continuous or Burst mode select PAUSE to terminate the current flow measurement but to leave the average flow mea surement and previous flow measurement on the s...

Page 19: ...igured table The data can be exported to se lectable Microsoft office environment To run DryCal Pro you must have Windows XP or 7 Microsoft Excel 2003 and up and a RS232 port or if your PC does not ha...

Page 20: ...17025 accredited laboratory such as our Butler NJ facil ity to ensure its measurement integrity For the ultimate in Defender maintenance and to take advantage of any available software and mechanical...

Page 21: ...by telephone at 973 492 8400 Our web site address is www mesalabs com Note Mesa Labs will not evaluate or service your instrument without an RMA number If we find the issues you have identified are ap...

Page 22: ...o its AC power adapter charger while in storage If your Defender cannot be attached to its AC power adapter charger while in storage please do the following Fully charge it before extended storage If...

Page 23: ...nits cc min mL min L min cf min Standardized Flow Units scc min smL min sL min scf min Pressure Units mmHg PSI kPa Temperature Units C F Basics Dimensions H x W x D 5 5 x 6 x 3 in 140 x 150 x 75 mm We...

Page 24: ...or Pressure Operating Pressure Absolute 15 PSI Operating Temperature 0 50 C Ambient Humidity 0 70 non condensing Storage Temperature 0 70 C Warranty 1 year battery 6 months DryCal Pro Software DryCal...

Page 25: ...h the following Default settings from the factory Reading Type Vol Number in Average 10 Time Between 0 Flow Units cc min Pressure Units mmHg Temperature Units C Standardizing Temp 21 1 C Measurement M...

Page 26: ...ories Inc shall have no liability to repair or replace any Mesa Laboratories Inc product 1 That has been damaged following sale including but not limited to damage resulting from improper electrical v...

Page 27: ...ender for pres sure sources or to the suction port for verifying suction pumps The unused port should be at atmospheric pressure with any cap or plug removed If you are calibrating a gas that requires...

Page 28: ...dead volume Dead volume is the gas volume between a flow generator and the instrument taking the measurement Since gas is compressible this gas can act as a spring between the flow source and the meas...

Page 29: ...ou We strive to provide the closest NIST traceable legal defensibility of any flow calibration equipment manufacturer and we actively maintain our NV LAP NIST ISO 17025 laboratory accreditation in ord...

Reviews: