2.7 Measurements
Taking gas flow readings
To maintain the best possible accuracy and minimize thermal effects, Mesa recommends
fully charging your battery before taking measurements. If this is not possible, we recom-
mend disconnecting your Defender from its AC power adapter/charger while taking flow
measurements — or to run gas through your Defender for 10 minutes before starting the
flow measurement.
First steps
Press the power button.
• Press the
On/Off
button for 1 second to turn on your Defender.
• When first turned on, your Defender displays an opening screen showing the product
name, model number and flow range.
• Press the
On/Off
button for 3 seconds to turn your Defender off.
Connect device to be calibrated to the appropriate Defender port. Use ¼ inch diameter
tubing.
• Connect to outlet at top (suction fitting) when a device draws air (such as sampler).
• Connect to inlet at bottom (pressure fitting) when a device pushes air.
• Do not cap the unused port on the Defender.
• Select the reading type to
Vol
or
Std
. Set ‘
Std To
’ to the desired standardizing tem
-
perature
• Choose the measurement type, Single, Burst, or Continuous, then press enter.