en
en
8
|
|
9
Setting time and date
When?
1. With a new instrument.
2. With the change of summer/winter time.
3. Check after replacement of batteries.
How?
1. Switch off the instrument by pressing the ON/OFF button.
2.
Press the MEM button and simultaneously press the ON/OFF key.
The time and date display blinks.
3.
To select the international or American date press the START button.
4. Press the TEST button to access hour, minute, day, and month.
5.
Press the START button to adjust the time (hour, minute) and the date
(day, month).
6.
End the procedure by pressing the MEM or the ON/OFF key.
The set values are automatically saved, the clock activated.
Initial calibration
1. Keep the instrument at room temperature for at least 30 minutes.
2. Switch the instrument on and insert the bar code of the recalibration set
(supplied) into the bar-code scanner. CAL appears in the display.
3. Insert the calibration strip of the recalibration set into the strip adapter as
shown in the diagram, and ensure that it is correctly positioned. Take utmost
care to ensure that the calibration strip is clean.
4. Press the START key. The instrument is recalibrated, the CAL message disappears
from the display, and the instrument switches off automatically.
5. Store the recalibration set in the supplied packaging materials, taking care to
protect it from light.
Important
When the batteries are low (display ) the instrument will not accept the
bar-code strip for calibration. Change the batteries first.
If the calibration strip is not inserted into the measurement chamber, an acoustic
signal sounds after the START button is pressed. The blinking CAL symbol in the
display prompts you to insert the calibration strip.
START
MEM
ON/OFF
TEST
2
START
MEM
ON/OFF
TEST
3
START
MEM
ON/OFF
TEST
4
START
MEM
ON/OFF
TEST
5