Operating Instructions DPG 2400 Series
29
Digital Pressure Gauge
DPG 2400 Series
No
UT?
#XUT?<cr><lf>
X<sp>UT<sp>n,
<sp>n,<sp>n,...
<cr><lf>
Returns the list of programmed
pressure units.
No
UT
#XUT<sp>n<sp>
N<cr><lf>
R
Loads the programmable units
list. The first number is the se-
quential position in the list from
0 to 7 and the second number
is the unit code as according to
Table 7.2.7.
No
ZC?
#XZC?<cr><lf>
X<sp>ZC<sp>nn
nnnnn
Returns the zero offset.
Yes
ZC
#XZC<sp>nnnnnn
<cr><lf>
R
Sets the zero offset.
7.2.6 User Programmable Pressure Units
The remote command
UT can be used to modify the default pressure units available on
the units display screen. The user can select from one to eight units to be displayed by
sending a sequence of “UT” commands. The syntax is as follows:
#*UT<sp>n<sp>n<cr><lf> where the first “n” is a value from 0 to 7 and represents the
units sequential position in the displayed list. The second “n” represents the unit code
from Table 7.2.7. If the first (0) unit in the sequence is assigned a value of “0” then the
displayed units will default to all the available units from the factory. To display less than
eight units, a “0” is assigned to the next unit in the sequence.
An example of the use of the “UT” command:
#*UT 0 1<cr><lf> - assigns the first unit displayed as PSI
#*UT 1 2<cr><lf> - the next unit displayed would be InHg 0C
#*UT 2 34<cr><lf> - the third unit displayed would be hPa
#*UT 3 0<cr><lf> - terminates the displayed units to this point
#*UT 0 0<cr><lf> - allows all factory units to be displayed and used
The “SAVE” command must be sent to make changes permanent.