page 15
10 Cooling system (special equipment)
If the waterbath is fitted with a cooling device
for quicker cooling of the bath liquid, the “cold
water supply“ has to be connected by a hose
to a cold water supply line. The “warm water
discharge“ has to be run by a hose to a drain.
Use temperature resistant material for the
tubing.
Please note, that the overflowing water may be up to 100ºC hot!
The overflow must be connected with tubing to an appropriate container or sink. Make sure, that the
tubing cannot be clogged or bended, and that it continually runs downhill.
11 Cleaning and maintenance
By regular cleaning of the easy to clean the tank, residues are
avoided which at continuous influence can impair the outfit and
function of the waterbath. Before using any other method for
cleaning or detoxicating than recommended by the manufacturer,
the user has to contact the manufacturer in order to make sure
that this chosen method will not destroy the appliance.
The unit can be cleaned with commercial stainless steel cleaning agent. It is important to ensure that no
rusting objects come into contact with the stainless steel bath tank or the stainless steel housing. Rust
sediments lead to contamination of the stainless steel tank or external casing.
If rust stains caused by contamination occur on the tank surface, the affected areas must be cleaned and
polished immediately.
Limeresidues can be removed by vinegar or citron-solutions.
On units with gable cover we recommend that the hinge bolts are oiled from time to time if the bath is
used frequently.
cold water
supply
warm water
discharge
BASIC