4.3.1 Temperature setpoint
The bath starts to heat up immediately
to the set temperature.
Adjustment range:
waterbaths:
10
°C to
95
°C or ca.
100
°C with
activation of the boiling mode
oilbaths:
10
°C to
200
°C
Setting and indication accuracy:
water-/oilbaths:
0.1
°C
Rotate the push/turn control until the
°C
symbol flashes.
The setpoint temperature can then be selected with the SET key depressed, as
described in Section 4.3. After releasing the SET key the display briefly flashes the
setpoint. The display then changes to the actual temperature and the controller
starts to control to the selected setpoint temperature.
During heating the symbol flashes in proportion to the actual heater power.
Waterbaths only:
In order to reach water temperatures above 95°C (boiling point), waterbaths are
provided with the boiling mode. If this is selected, the heating is switched on
permanently.
Activating the boiling mode by rotating the push/turn control beyond the
maximum value until the display shows
CCC
The setting “boiling mode“ is not stored permanently. After the bath is switched
off and on again, the controller shows again the previously selected setpoint..
°C
°C
4.3.2 Delayed switch-on
The bath starts to heat up to the previously
selected temperature only after the time of
the switch-on delay has elapsed.
Adjustment range:
1
min bis
999
hrs
Setting accuracy:
<100 hrs:
1
min
≥
100 hrs:
1
hr
Indication accuracy:
<10 hrs:
1
min
≥
10 hrs:
1
hr
Rotate the push/turn control until the (delay) and the symbol flashes.
The duration of the delayed switch-on can then be set with the SET key depressed
as described in Section 4.3. After the SET key has been released the bath briefly
flashes the setting of the switch-on delay.
The switch-on delay is then activated and the display alternates between the
actual temperature and the running time of the switch-on delay. The time is
shown with a negative sign and runs down. In this way it is possible to determine
at any time how much longer the bath waits until it begins to heat up.
If no delayed switch-on is required, it can be de-activated by the setting
OFF
.
°C
°C
°C
*
90°C
*
delay 12h
90°C