Menu functions
1. Activate the
SETUP
display and select
Alarm temp
with the turn control.
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
Slid i
i
2. Accept the selection by pressing the con-
firmation key. The adjustment options are
automatically highlighted.
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
3. With the turn control, select the desired
setting – in this example TWB.
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
Slid i
it
4. Save the setting by pressing the confir-
mation key.
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
Slid i
it
5. Select
Max Alarm
with the turn control.
SETUP
Max Alarm
190 °C
2/2
6. Accept the selection by pressing the
confirmation key. The current settings are
automatically highlighted.
SETUP
Max Alarm
190
°C
2/2
7. With the turn control, select the desired
new trigger temperature - in this example
160 °C.
SETUP
Max Alarm
160
°C
2/2
8. Save the setting by pressing the confir-
mation key. The electronic temperature
monitoring system will now be triggered
when the actual temperature reaches 160
°C.
SETUP
Max Alarm
160 °C
2/2