36
D33363 | Date 06/2016
Menu functions
All other settings can be made accordingly. The settings possible are described in the follow-
ing sections.
If no new values are entered or confirmed for approx. 30 seconds, the appliance automati-
cally returns to the main menu and restores the former values.
7.3
Setup
In the
SETUP
display, you can set the following parameters:
Ź
the
IP address
and
Subnet mask
of the appliance's Ethernet interface (for connection to
k
a network)
Ź
the
Unit
of the temperature display (°C or °F, see page 37)
Ź
the type of temperature monitoring (TWW or TWB,
Alarm Temp,
only if equipped with
a second temperature sensor, option A6) and the trigger temperature of the monitoring
function (
Max alarm
, see page 38)
Ź
how the digital backwards counter with target time setting works (
Timer mode
, see page
39)
Ź
the type of the slide-in unit (Grid or Shelf, see page 39)
Ź
Remote control (see page 39)
Ź
Gateway (see page 40)
If the SETUP menu contains more entries than can be
displayed, this is indicated by the display “1/2”. This
means that there is a second „page“ of entries.
To display the hidden entries, use the turn control
to scroll beyond the lowest entry. The page display
changes to “2/2”.
7.3.1 IP address and subnet mask
If you want to operate one ore more appliances in a network, each appliance must have its
own unique IP address for identification. By default, each appliance is delivered with the IP
address 192.168.100.100.
LAN 1: 192.168.1.233
192.168.1.216
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241
Editor
Programmname
Simulation
Protokoll
-
+
-
+
INP 250
Test 01
180
.0°C
i
HPP 250
Labor
i
37
.0°C
44
.4 %rh
15
15
STAND BY
Programm
n
e
g
n
u
l
l
e
t
s
n
i
E
l
l
o
k
o
t
o
r
P
t
ä
r
e
G
e
f
l
i
H
n
e
k
c
u
r
D
AtmoCONTROL
Fig. 22
Operation of several appliances in a network (schematic example)
Setup
Unit
IP adress
255.145.136.2
Subnet mask
255.255.0.0
°C
F
1/2