2 Installation requirements
8
On-site requirements
Mains supply
Implement the following safety measures when dealing with the cable and power plug:
u
Never damage or alter the power plug or cable.
u
Never bend or twist the power cable.
u
Never remove the plug by pulling on the power cable. Always take a grip on the plug.
u
Never place any heavy objects on the power cable.
u
Never run the power cable over areas in which it could become trapped (e.g. doors or windows).
u
Never lead the cable along a source of heat.
u
Never use any nails, paper fasteners or similar objects to fix the cable.
u
Should the power plug or cable suffer damage, switch off the device. The power cable or plug
should only be replaced by authorised technicians.
u
The mains socket must be freely accessible after installation so that the steam sterilizer can be
disconnected from the electricity supply at any time.
On-site requirements of the mains connection
Property
Electricity supply
220-240 V, 50/60 Hz
Max. voltage range
207-253 V
Building fuses
separate power circuit with 16 A fuse, 30 mA RCD protection (to guarantee
continued practice operation during steam sterilizer malfunction)
Length of power cable*
)
1.35 m
Other
additional socket for the MELAprint 42/44 log printer etc.
*
)
Comply with the specifications in the connection diagram.
Water connection
Feed water
Wastewater
Connection in the
practice
Manual filling via the internal storage
tank
Optional: to a water treatment unit,
e.g. MELAdem
Manual emptying via the internal
storage tank.
Optional: automatically via the one-
way outlet with the MELAG upgrade
kit for the tank outlet
Wall outflow, nominal width DN 40 or
to a siphon (flush outflow)
Installation height
--
min. 30 cm under the steam sterilizer
Max. water temperature
35 °C
70 °C
Recommended flow
pressure
1.5 bar at 3 l/min
--
Min. water pressure
(static)
corresponding to the water treatment
unit
1)
--
Max. water pressure
(static)
10 bar
--
Max. water
consumption
2)
approx. 700 ml
--
Water quality
Distilled or demineralised water in
accordance with EN 13060, Appendix
C
--
1)
Optional when using a water treatment unit
2)
In the Prion program with porous full load.
Summary of Contents for Vacuklav 23 B+
Page 2: ......