3 Setup and installation
17
PLEASE NOTE
Before converting from the standard version to a water treatment unit, you must first
empty the feed water side (right-hand chamber) of the internal storage tank. With
automatic one-way outlet, the wastewater side (left-hand chamber) must be emptied as
well.
Attaching the safety combination in accordance with EN 1717
1.
Unscrew the back panel of the steam sterilizer.
2.
Mount the safety combination (art. no. ME82375).
3.
Connect the water inflow hose to the upper connection of the safety combination.
4.
Connect the water inlet hose to the MELAdem 47 reverse osmosis unit to the lower connection of the
safety combination.
5.
Screw the back panel back on. Guide the hoses through the opening provided at the bottom edge of
the back panel.
Aligning the steam sterilizer
To ensure fault-free operation, the steam sterilizer must be set up horizontally with the aid of a spirit level
placed on the chamber seal face. Then extend the fore device feet by five (Vacuklav 23 B+) or three
(Vacuklav 31 B+) revolutions to effect a slight rearwards slope of the device.
Test runs
Vacuum test with a cold sterilization chamber
Perform a vacuum test with an empty, cold sterilization chamber and record the outcome in accordance
with record of installation and setup.
Universal-Program
If the vacuum test was successful, perform a Universal-Program with 1.5 kg load (instruments) and record
the outcome in accordance with record of installation and setup.
Instructing the users
Explain all the user-typical features for the documentation and setting combinations for the operator.
Hand over the manufacturer’s inspection report. The declaration of conformity regarding the pressure
equipment directive and the Medical Devices Directive are included in the manufacturer's inspection re-
port.