2 Installation requirements
7
The steam sterilizer works with a cooler on the rear of the device. The function and life-span of the steam
sterilizer can be compromised if heat dissipation via the cooler is restricted in any way. The steam sterilizer
may only be installed if sufficient air circulation can be guaranteed.
Additional space requirement for the feed water supply
Additional space is required for a storage container or a water treatment unit. It is also necessary to
guarantee free access to the hoses and cables leading from the steam sterilizer to the water treatment
unit.
Space requirements
MELAdem 40
MELAdem 47
Water treatment unit
Pressure tank
Width
32 cm
40 cm
--
Height
35 cm
46 cm
40 cm
Depth
16 cm
18 cm
--
Diameter
--
--
approx. 28 cm
Requirements for the installation of a device
If the installation of the device is mandatory, one of the following measures must be implemented:
1.
It must be possible to pull out the device for operation (pos. a).
2.
There must be an exhaust air shaft in the rear area of the installation space that discharges the warm
air upwards (pos. b).
3.
There must be an exhaust air shaft in the rear area of the installation space that discharges the warm
air to the rear (pos. c).
4-5 cm
a
b
c
Summary of Contents for Vacuklav 23 B+