background image

User Manual

Sterilizer 75

Dry-heat sterilizer

EN

Dear customer,

We thank you for your confidence demonstrated by the purchase of this MELAG product. As an owner-run and operated
family concern founded in 1951, we have a long history of successful specialization in hygiene products for practice-
based use. Our focus on innovation, quality and the highest standards of operational reliability has established MELAG
as the world’s leading manufacturer in the instrument treatment and hygiene field.

You, our customer are justified in your demand for the best products, quality and reliability. Providing 

"competence in

hygiene"

 and 

"Quality – made in Germany"

, we guarantee that these demands will be met. Our certified quality man-

agement system is subject to close monitoring: one instrument to this end is our annual multi-day audit conducted in ac-
cordance with ISO 13485. This guarantees that all MELAG products are manufactured and tested in accordance with
strict quality criteria.

The MELAG management and team.

Summary of Contents for Sterilizer 75

Page 1: ...shed MELAG as the world s leading manufacturer in the instrument treatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Providing compe...

Page 2: ......

Page 3: ...air sterilizer on and off again 11 Opening and closing the door 12 Changing the mount 12 6 Sterilization 13 Preparing the sterilization material 13 Loading the dry heat sterilizer 13 Information regar...

Page 4: ...to the device n The power cable may not be replaced by a cable determined to be insufficient Danger of short circuit n Liquids may not be permitted to reach the interior of the device This could resu...

Page 5: ...perating Manual near to the device It represents a component of the product Formatting rules Example Explanation see Chapter 2 Reference to another text section within this document Symbols used Symbo...

Page 6: ...the danger of burns In affixing the CE mark the manufacturer declares that this product fulfils the basic requirements of the EMC and low voltage directive The device may not be disposed as domestic...

Page 7: ...ack Standard tray jack Standard tray cassette unperforated Intended use This hot air sterilizer is intended for use in the cosmetics sector the cosmetic chiropody and the veterinary sector It was desi...

Page 8: ...amp heating 3 Thermometer 4 Indicator lamp network 5 On Off switch and time setting 6 Door handle 7 Door opens forwards 8 Device foot 9 11 10 Fig 2 Rear view of the device 9 Fan 10 Connection for the...

Page 9: ...loading versions with devices with trays and standard tray cassettes are possible in the following siz es Tray WxHxD 19 x 2 x 29 cm Standard tray cassette WxHxD 19 x 4 x 29 cm Mounting 1 Mount 1 belo...

Page 10: ...clearance of 10 cm to other devices and walls especially flammable parts u Ensure sufficient ventilation Ensure sufficient clearance to the top for free removal of the warm air The device may not be u...

Page 11: ...wise The control lights network and heating will remain illuminated until the time expires The hot air sterilizer can be set to continuous operation In this case the hot air sterilizer must be switche...

Page 12: ...cedure has started this could result in cooling and the object not being steri lized for a sufficient time 1 Move the On Off switch to the 0 position to switch off 2 Open the door forwards to open 3 M...

Page 13: ...tion material together in blocks as this will prevent heat equalization u Do not place the sterilization material on cellulose as this can result in a heat build up u Always pack the sterilization mat...

Page 14: ...zation times table Selecting the sterilization times Do not exceed the specified loading quantities Use only trays and cartridges made of aluminium When needing to pack with foil use aluminium foil Ta...

Page 15: ...a specified below for the storage of sterilized equip ment u Comply with the maximum storage duration in accordance with the packaging type u Do not store the sterilized equipment in the decontaminati...

Page 16: ...ng bio indicators thermocouples or maximum thermometers Comply with the valid regional legal specifications When performing the spore test using biological indicators the spore packets are to be clamp...

Page 17: ...o remove the impurities with this method Only if the chamber mount or chamber seal face has persistent soiling should you use a mild stainless steel cleaning agent with a pH value between 5 and 8 To c...

Page 18: ...temperature has been set too low The temperature controller rotary knob was unintentionally turned or mal adjusted during unpacking or incorrect handling e g during cleaning Set a higher temperature T...

Page 19: ...ase observe the maximum load 2 kg including trays Maximum mains voltage supply variations 10 Maximum voltage range 207 253 V Relative humidity up to 31 C max 80 up to 40 C max 50 Degree of protection...

Page 20: ...cessories 20 11 Accessories Table 2 Accessories Designation of article art no Mounting 1 59030 Mounting 2 59040 Tray 02075 Standard tray cassette 00287 Tray jack 28890 Jack for standard tray cassette...

Page 21: ...e sterilization material to the required temperature of 180 C Sterilization material Unsterile sterilizable material which is still to be sterilized Sterilized equipment Also referred to as a batch a...

Page 22: ...0829 Berlin Germany Email info melag com Web www melag com Original instructions Responsible for content MELAG Medizintechnik GmbH Co KG We reserve the right to technical alterations BA_Hei luftsteril...

Reviews: