6 Sterilization

14

Information regarding routine operation

The sterilization temperature must amount to 180 °C before the start of every sterilization procedure. The
equilibration time must be taken into account when setting the sterilization time. The pure elimination time
amounts to 30 min at 180 °C.

1.

Wait 15 min after the thermometer has reached 180 °C.

2.

Set the required sterilization time. Comply with the information in the sterilization times table.

3.

Check whether the mechanical air movement is functioning correctly via the ventilator noise.

Setting the temperature

Set the temperature for the hot air sterilization as follows:

1.

Turn the temperature controller to 180 °C.

2.

Wait until the thermometer displays 180 °C and allow the device to heat up for 15 min.

Ê

During this time, the hot air sterilizer heats up the complete sterilization chamber.

3.

Set the sterilization time via the On/Off switch. Comply with the information in the sterilization times
table.

Selecting the sterilization times

Do not exceed the specified loading quantities. Use only trays and cartridges made of aluminium. When
needing to pack with foil, use aluminium foil.

Load quantity

Packaging

Pre-heating time

Operating time

max. 1 kg inc. 2 trays

unwrapped

15 min

60 min

max. 4 kg inc. 6 trays

unwrapped

15 min

75 min

max. 1 kg inc. 1 standard tray cassette

wrapped

15 min

75 min

max. 4 kg inc. 3 standard tray cassettes

wrapped

15 min

120 min

Removing the sterilized equipment

CAUTION

Danger of burns from hot metal surfaces

n

Allow the device to cool sufficiently before opening.

n

Do not touch any hot metal parts.

CAUTION

Unsterile instruments resulting from damaged or burst packaging. This endangers the
health of your patients and practice team.

n

Should the packaging be damaged or have burst after sterilization, re-pack the sterilization
material and re-sterilize it.

Comply with the following when removing the sterilized equipment:

u

Use a tray lifter to remove the tray.

u

Never touch the sterilized equipment, the device interior or the inside of the door with unprotected
hands. The components are hot.

Summary of Contents for Sterilisator 205

Page 1: ...lished MELAG as the world s leading manufacturer in the instrument treatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Providing com...

Page 2: ......

Page 3: ...ir sterilizer on and off again 11 Opening and closing the door 12 Replacing the mount 12 6 Sterilization 13 Preparing the sterilization material 13 Loading the dry heat sterilizer 13 Information regar...

Page 4: ...legible damaged or lost obtain a new copy from MELAG State the device type and your address in an e mail The device type is specified on the type plate on the rear of the device Formatting rules Examp...

Page 5: ...quids may not be permitted to reach the interior of the device This could result in an electrical shock or short circuiting Preparation and sterilization n Do not place the objects to be sterilized on...

Page 6: ...ount 2 depending on the order Intended use This hot air sterilizer is intended for use in the cosmetics sector the cosmetic chiropody and the veterinary sector It was designed for the sterilization of...

Page 7: ...controller 3 Control lamp heater 4 Thermometer 5 Control lamp mains 6 On Off switch and time setting 7 Door handle 8 Door opens forwards 11 9 10 Fig 2 Rear view of the device 9 Fan 10 Type plate 11 Co...

Page 8: ...s symbol indicates that the marked area becomes hot during operation Contact with it during or shortly after operation can pose the danger of burns In affixing the CE mark the manufacturer declares th...

Page 9: ...l include a tray lifter The following loading versions are possible in devices with trays and standard tray cassettes of the follow ing sizes Tray W x H x D 19 x 2 x 36 cm Standard tray cassette W x H...

Page 10: ...alls especially flammable objects u Ensure sufficient ventilation Ensure sufficient clearance to the top for free removal of the warm air The device may not be used as an installation device and not b...

Page 11: ...control lamp will remain illuminated the control lamp for the heater will switch on and off until the time has elapsed The hot air sterilizer can be set to continuous operation In this case the hot a...

Page 12: ...sult in cooling and the objects not being sterilized for a sufficient time 1 Move the On Off switch to the 0 position to switch off 2 Move the door forwards to open 3 Move the door backwards to close...

Page 13: ...l on cellulose as this can result in a heat build up u Unwrapped sterilized equipment looses its sterility upon contact with the ambient air If you plan to store your instruments in a sterile fashion...

Page 14: ...erilization times table Selecting the sterilization times Do not exceed the specified loading quantities Use only trays and cartridges made of aluminium When needing to pack with foil use aluminium fo...

Page 15: ...he criteria specified below for the storage of sterilized equip ment u Comply with the maximum storage duration in accordance with the packaging type u Do not store the sterilized equipment in the rep...

Page 16: ...ng bio indicators thermocouples or maximum thermometers u Comply with the valid regional legal specifications u When performing the spore test using biological indicators the spore packets are to be c...

Page 17: ...d components When cleaning the sterilization chamber the mount for the load and the door seal comply with the following u Switch off the device before cleaning and disconnect the power plug from the s...

Page 18: ...set too low The temperature controller rotary knob was turned or set incorrectly when unpacking the device or because of incorrect handling e g during cleaning Set a higher temperature The temperature...

Page 19: ...220 240 V 50 60 Hz 110 V 50 60 Hz Maximum voltage range 207 253 V Maximum mains voltage supply variations 10 Electrical power 1100 W Device fuse 2 x 6 3 A T 12 5 A T 110 V Ambient conditions Installat...

Page 20: ...u can obtain the specified articles and an overview of further accessories from your stockist Article Art no Mount 1 60030 Mount 2 60040 Tray 02000 Standard tray cassette 00287 Tray lifter 28890 Lifte...

Page 21: ...Sterilisiergutes die geforderte Temperatur von 180 C zu erreichen DIN 58953 Standard sterilization sterile equipment supply Sterilization material Unsterile sterilizable material which is still to be...

Page 22: ...10 10829 Berlin Germany Email info melag com Web www melag com Original instructions Responsible for content MELAG Medizintechnik GmbH Co KG We reserve the right to technical alterations BA_Sterilisat...

Reviews: