4 Adapter for the Carebox
11
Adapter installation
Comply with the following for safe handling:
n
The Carebox Blue is used for the reprocessing of instruments that must be maintained with maintenance oil.
n
The Carebox Green is used for the reprocessing of instruments that must not be maintained with maintenance oil.
n
Use only original accessories.
The following must be fulfilled or present:
ü
The adapters for transmission instruments, hollow-body instruments or unused connections.
ü
A new sealing washer, a new O-ring and screws for each adapter.
ü
A TX6 screwdriver.
1.
Loosen and remove the screws of the existing adapter.
2.
Remove the adapter.
3.
Remove the sealing washer and any residue from the seal.
Be careful not to scratch the seal face of the Carebox.
4.
Insert the new sealing washer by aligning it with the recess and hole
pattern.
5.
Insert the new O-ring into the adapter.
6.
Insert the new adapter by aligning it with the recess.