Working with the Microscope
Please read the following instructions to achieve the best microscope observing results.
Setting the illumination
For optimum effect in contrast and resolution one should follow the below procedure:
Place a specimen on the object stage and focus using the 4x objective, with a fully opened
iris diaphragm.
•
Turn the condenser in the highest position.
•
Close the iris diaphragm, until it is just visible on the edge of the field of view.
The microscope is properly set for use with the 4x objective. For each other magnification
this procedure should be repeated to ensure the best balance between contrast and
resolution.
Caution:
The maximum light intensity when using the 4x and 10x can damage the eyes!
When changing slides always start with the 4x objective again.
Safety device
To prevent damage to the objective lens, or breaking the slide, all types are equipped with a
pre-fixed safety device (F).
It is recommended to use slides of 1.0 – 1.2 mm thickness in combination with cover glasses
of 0.13 mm or 0.17 mm thickness.