MT-10
Series
Compound Microscope
INSTRUCTION MANUAL
The total magnification can be calculated by multiplying the magnification of the eyepiece
with the magnification of the objective. The magnifications are displayed in the table below:
Eyepiece
Objective
Magnification
10x
4x
40x
10x
10x
100x
10x
40x
400x
10x
60x
600x
10x
100x
1000x
Object stage
On the
MT-10
series the slide is placed beneath the object clips. Focusing the specimen is
done by using the coaxial coarse- and fine adjustment knobs (H).
Coarse- and fine adjustment
The coarse- and fine adjustment knobs (H) for the height adjustment of the object stage
are mounted together on one axis (co-axial). On both fine adjustment knobs there is a
graduation with intervals of 0.002 mm. This can be used to measure depths in a specimen.
Abbe condenser with iris diaphragm
Beneath the object stage an Abbe condenser (L) N.A. 12.5 is mounted. The condenser can be
adjusted in height by means of turning or by a handle (polarization models). With this one
can focus the light on the specimen to optimise contrast-resolution ratio. The condenser is
factory pre-centered.
The iris diaphragm with filter holder is mounted beneath the condenser. The light intensity
can be adjusted by changing the flexible opening.
Illumination MT-10 series
In order to have good color rendering against the Michel-Levy chart, the polarization
versions are standard equipped with a halogen illumination.
All other models have a LED illumination
The cable goes to the mains socket at the backside of the microsocpe.