background image

Balance Control

Without Balance Control (see point 3 in the User Controls and
Display section) the transmitted pulse would be visible at the
beginning of the trace, swamping any reflections within the
pulse length (the dead zone). The balancing circuit attempts to
match the characteristic impedance of the cable under test to
produce an equivalent pulse. Subtracting this equivalent pulse
from the transmitted pulse effectively removes the dead zone
and allows cable features much closer in to be detected.
Alternatively, using the [L1-L2] & [OFF] mode, where L2 is
connected to a known good length of the cable under test, L2 is
used instead of the balancing circuit to automatically null the
transmitted pulse.

NOTE: In many cases, it will be impossible to completely null 

the transmitted pulse.

Velocity Factor

The velocity factor is the scalar that is used to convert the
measured time interval into an actual length of cable. It can be
displayed in one of two ways: a ratio of the transmitted pulse
speed to the speed of light, or as a distance per microsecond.
When it is displayed as the distance per µs (either m/µs or
ft/µs) the velocity factor will be indicated as half the speed of
the pulse in the cable. This is because the pulse in fact has to
go along the cable to the cable feature and back again which is
twice the distance to the feature.

The table of velocity factors in the HELP pages of the
instrument is a rough guide and in practice, the settings are

subject to many variable factors. If the exact length of a piece
of cable of the same type as the C.U.T is known and the
reflection from the cable end is visible then a more accurate
value can be determined:
1. Locate the reflection caused by the end of the known length 

of cable with the instrument set on the shortest possible     
range to see the end of the cable.

2. Locate the start of this reflection as described in the  

Operation section of this manual.

3. Adjust the velocity factor until the correct cable length is 

shown.

The measurement of the distance to the fault can now be made
with more confidence that the measurement will be correct. The
ability of the instrument to accurately measure the distance to a
cable feature relies on the velocity factor being correct, any
percentage errors in the velocity factor are directly proportional
to distance measurement errors. Hence, the TDR2000 uses the
velocity factor to three decimal places to reduce any errors.

Pulse Widths

The TDR2000 pulse widths range from 20ns to 16µs to
overcome signal attenuation and enable the instrument to see
further down a length of cable. In distance terms for the size of
the transmitted pulse, this represents a transmitted pulse from
as small as 4.0m to 3199m! (This assumes a velocity factor of
0.667.) Without Balance Control, this would be an enormous
dead zone, but with the instrument correctly balanced, faults
can be seen well within the pulse width. 

9

Instrument Features:

Summary of Contents for TDR2000

Page 1: ...User Guide MEGGER Time Domain Reflectometer MEGGER CFL535E TDR2000 Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 Established 1981...

Page 2: ...Techniques to Improve Accuracy 13 Care and Maintenance 13 Specification 14 Repair and Warranty 16 MeterCenter 800 230 6008 www MeterCenter com Caution Refer to accompanying notes Equipment protected t...

Page 3: ...y the manufacturer to reduce the hazard the user must assume responsibility for ensuring his or her own safety The instrument should not be used if any part of it is damaged Test leads probes and croc...

Page 4: ...will occur the cable appearing endless Open or Short circuits will reflect all the pulse energy and the TDR will not see the cable beyond that fault As a pulse is transmitted down a cable the size an...

Page 5: ...ransfer of the waveform data to a computer for analysis and storage for future reference Other setting options include changing the distance units between metres and feet changing the propagation velo...

Page 6: ...e under test When balanced the transmitted pulse can be nulled out and cable features close to the leads can be detected 4 L2 Sockets The sockets are designed to accept the leads supplied with the ins...

Page 7: ...nt screen to an attached EPSON compatible printer as a screen dump If pressed with the shift key then the instrument will go to computer controlled operation where a computer running the appropriate d...

Page 8: ...able feature e g an open or closed circuit move the cursor to the very beginning of the reflection To locate the start of the reflection more accurately press the DISPLAY key to zoom in around the cur...

Page 9: ...ty factors in the HELP pages of the instrument is a rough guide and in practice the settings are subject to many variable factors If the exact length of a piece of cable of the same type as the C U T...

Page 10: ...the default pulse width can not discern between the user can override the default by entering the CONFIG Options menu Memory Features The TDR2000 has 15 memory locations which can be used to store tra...

Page 11: ...mines which waveforms will be acquired for display as traces 1 2 Though the MODE key can select between the four basic modes of operation see the introduction the user has more ability to select modes...

Page 12: ...velocity factor can be displayed as a ratio of pulse speed to the speed of light or as a distance per microsecond This option selects the display type Please refer to the Instrument Feature section fo...

Page 13: ...the cable from both ends Particularly in the case of open circuit faults the true end of the cable is not visible Thus it is harder to estimate whether the answer obtained is realistic If the measure...

Page 14: ...electable 50m range 20ns 40ns 60ns 80ns 100ns 100m range 20ns 50ns 800ns 100ns 140ns 200m ranges 20ns 60ns 100ns 140ns 200ns 400m range 40ns 80ns 160ns 200ns 400ns 1km range 80ns 160ns 260ns 500ns 1 s...

Page 15: ...ed then an additional blocking filter must be used EMC Complies with Electromagnetic Compatibility Specifications Light industrial BS EN50081 1 1992 BS EN50082 1 1992 Mechanical The instrument is desi...

Page 16: ...oad Valley Forge Corporate Centre Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8579 Fax 44 0 1304 207342 Fax 1 610 643 8625 For sales...

Reviews: