Resistance range (MIT 320 and MIT330 only)

(can be used for diode testing)

Measuring range: 

  

10 Ω - 1 MΩ 

Open circuit voltage:

  

5 V

Short circuit current:

  

1.5 mA,

Accuracy (at 20°C): 

 

±5% ±2 digits up to 100 kΩ

Measuring range (MIT310A only): 

0 - 2 kΩ (refer to continuity)

Voltage range 

Measuring range:

   

0 - 600 V a.c. (50/60 Hz) or d.c.

Accuracy (at 20° C)

MIT300, 310, 320, 330:

 

d.c. or a.c. (50/60 Hz): ±1%,  

   ±2 

digits

MIT310A:  

 

±2.5% of scale length for 50/60 Hz

Temperature coefficient:

 

<0,1% per °C on all ranges.

Default voltmeter 

all test modes except off 

MIT310, 320, 330 voltmeter:

>25 V a.c. or d.c. is applied display will    

 

 

 

operate as a voltmeter

M300:

 

 

 

Beeps and flash ‘V’ on the display

M310A:

   

 

Warning LED lights when [TEST] button is  

 

 

 

pressed. Operates as voltmeter when test  

    button 

not 

pressed

Test inhibit:

 

 

If more than 50 volts is detected, testing   

   will 

be 

inhibited.

Autopower down:

  

Auto power down operates after 10  

 

 

 

 

minutes if left in standby mode. 

Environmental 

Operating range:

   

-10 °C to +60 °C

Operating humidity: 

93% R.H. at +40 °C max.

Storage range:

 

 

-25 °C to +70 °C

Environmental protection:

   IP54

Results storage and downloading

Capacity: 

  >1000 

results

Download : 

 

USB download lead 1.8m (supplied)

Download lead: 

 

Type A to Type B

Dimensions  

 

All units 203 x 148 x 78 mm

Fuses

Terminals: 

 

 

500 mA (F) 600 V, 32 x 6 mm ceramic    

 

 

 

HBC 50 kA minimum.

Safety

Meets the requirements of IEC61010-1 Cat III 600 V phase to earth. Refer 
to safety warnings supplied.

Automatic discharge

After an insulation test the item under test will be discharged automatically. 
Any voltage present will be indicated on the display so that the discharge 
can be monitored.

Power supply

Summary of Contents for MIT300

Page 1: ...M MIT300 Series Insulation and Continuity Testers USER MANUAL...

Page 2: ...of fused test leads when measuring voltage on high energy systems n Replacement fuses must be of the correct type and rating Failure to fit the correctly rated fuse may result in a safety hazard and...

Page 3: ...warnings 4 Introduction 2 General description 4 Case contents 5 LCD display 6 Front panel digital instruments 6 Connection panel 7 Lid open closure 7 Preparations for use all instruments 8 Batteries 8...

Page 4: ...1 to 20 n n Audible buzzer disable n n Lead null to 9 0 0 5 for MIT310A n n 0 5 n n Stored test lead null n n n n n Resistance 10 to 1M resistance range n n INTRODUCTION GENERAL DESCRIPTION 4 Thank yo...

Page 5: ...Certificate of test 1 x Calibration certificate 1 x CD containing user manual 1 x Printed quick start guide 1 x USB lead MIT330 only 1 x SP4F switched probe MIT320 and MIT330 only CASE CONTENTS 5 MIT...

Page 6: ...st lock Measurment ranges Analogue pointer Battery condition indicator Warning refer to user guide Multi fuction analogue range display Fuse blown indicator Continuity indicator Test lock warning Infi...

Page 7: ...panel tab 1 2 Fold away underneath instrument 2 3 and push into retaining slot 4 Connection Panel MIT300 Series test lead connections An interlock sliding panel prevents connection of unsafe test lea...

Page 8: ...nge all the time the TEST button is locked down Voltage warning MIT300 only If a voltage exists on the circuit under test the instrument will emit a warning beep If a voltage greater than 50 V exists...

Page 9: ...operator should refer to the user manual for further information LRange lock Displayed at any time the TEST button is locked in the on position zLead nulled Indicates that the test lead resistance ha...

Page 10: ...istance should be zeroed out This ensures any measurement does not include the resistance of the test leads 1 Connect the test leads to the instrument and firmly short both leads together 2 Wait for t...

Page 11: ...st check which looks for a resistance greater than 1 0 M and will buzz when found If a voltage is detected the continuity test will stop the instrument will not be damaged The continuity buzzer will t...

Page 12: ...test probes to the isolated circuit under test 3 Press and hold the TEST button allow the display to settle The reading will remain displayed for a few seconds after the push button is released 4 Rel...

Page 13: ...o OFF and the threshold value set to 1 00 M To adjust 1 Switch the instrument to SETUP 2 Repeatedly press the TEST button until the display shows RES 3 The initial insulation alarm level is set to 1 0...

Page 14: ...ion of the needle over the taper symbol will indicate battery condition Highest charge to the right Note NiMH or NiCAD rechargeable batteries show a lower charge than alkaline batteries and my not giv...

Page 15: ...ance Continuity test MIT310A Continuity range extends from 0 00 to 2 k Ranges are not auto ranging Manual selection of each range is necessary Test currents are 200 mA to 2 20 mA to 20 2 mA to 200 200...

Page 16: ...1 P2 Phase number each test can be stored under a particular phase P1 P2 or P3 Unique test number each test result is assigned a unique test number from 0 to 1999 This is automatic and cannot be adjus...

Page 17: ...ar graph shows the progress of the deletion 17 2 Select job reference number using keys then press NEXT Hold the key down to scroll quickly through the numbers 3 Select distribution board number b01 0...

Page 18: ...ay not give much warning before becoming exhausted To replace batteries Warning Do not switch the instrument on with the battery cover removed 1 Switch off the instrument and disconnect the instrument...

Page 19: ...he fuse symbol findicates a ruptured fuse The fuse is located behind the rear cover The rear cover must not be opened if test leads are connected To avoid the possibility of shock disconnect the batte...

Page 20: ...d IEC 364 Accuracy at 20 C MIT300 310 320 330 3 2 digits up to 10 M 5 up to 100 M 30 up to full scale MIT310A 2 5 of scale length or 30 of reading 200 k to 10 M Terminal characteristics Continuity ran...

Page 21: ...perates as voltmeter when test button not pressed Test inhibit If more than 50 volts is detected testing will be inhibited Autopower down Auto power down operates after 10 minutes if left in standby m...

Page 22: ...E M C In accordance with IEC61326 1 Operational uncertainties Refer to www megger com Order Code Included Accessories Test lead set red black with crocodile clips 1002 001 USB download lead MIT330 on...

Page 23: ...eturning your product to Megger UK and USA service centres 1 When an instrument requires recalibration or in the event of a repair being necessary a Returns Authorisation RA number must first be obtai...

Page 24: ...nce with Directive 2014 53 EU Other equipment manufactured by Megger Instruments Limited described in this user guide is in compliance with Directives 2014 30 EU and 2014 35 EU where they apply The fu...

Page 25: ...megger com Megger Pty Limited Unit 1 11 21 Underwood Road Homebush NSW 2140 T 61 0 2 9397 5900 F 61 0 2 9397 5911 E ausales megger com Megger Limited Unit 106 550 Alden Road Markham Ontario L3R6A8 Ca...

Reviews: