background image

Storing insulation test results:

 

Storing an insulation test result is exactly the same as a continuity test.  
However the circuit type options change to L-E, L-L, L-N, N-E

Downloading results to a PC:

1.  Connect the tester to the PC using the USB test lead.
2.  Set the tester range knob to [Snd]. 
3.  Run Megger Powersuite Professional or Megger Download Manager 

on  

 

the PC

4. Select the appropriate downloading instrument from the list
5. Select “download from tester”
6. The test data will automatically download its contents to the  

 

 PC.

A bar graph shows the status of the download.

Deleting test results

To delete the latest test result:
1.  Set the range knob to [dEL]. The display will flash “dEL”    

 

followed by the test number to be deleted.

2.  Press the OK button. The last test result will be will be deleted.

WARNING:

 This operation is not reversible. 

To delete all test data:

1.  Set the range knob to [dEL]. The display will flash “dEL”.
2.  Press the NEXT or LAST key. The display will flash “ALL”.
3. Press the OK button. All the test results will be deleted.
  A bar graph shows the progress of the deletion. 

17

2. Select job reference number using 

­↓

 keys then press NEXT.

  (Hold the key down to scroll quickly through the numbers)
3.  Select distribution board number (b01,02 etc) using 

­↓

 keys  

 

then press NEXT

4.  Select circuit number (c01,02 etc) using 

­↓

 keys then press NEXT

5. Select the circuit type, R12, RrN, Rr2, Rr1, R2 or R1 by pressing the 

­↓

    

keys, then press NEXT.

6. Select the phase using 

­↓

 keys then press NEXT.  The screen will  

 

display a unique test number , which is attached to that particular test. 

7. Press OK to save the result.

To store a subsequent result:

To save the next test under the same location job number, distribution 
board, circuit reference etc:

1. Make another measurement as described earlier and press STORE.
2.  The last Job number ** will be displayed. Press OK.
3.  The unique test number will be displayed. Press OK and the result is    

stored.

Note:

 The job number, distribution board, circuit and test type can all be 

changed before saving a result by selecting the function using the 

­↓

 keys.

To recall the last test result:

1. Set the range knob to RCL
2.  The last unique test number is displayed
3.  Press OK and the test result will be displayed.

Note:

 Only the last test result can be recalled to the display.

Summary of Contents for MIT300

Page 1: ...M MIT300 Series Insulation and Continuity Testers USER MANUAL...

Page 2: ...of fused test leads when measuring voltage on high energy systems n Replacement fuses must be of the correct type and rating Failure to fit the correctly rated fuse may result in a safety hazard and...

Page 3: ...warnings 4 Introduction 2 General description 4 Case contents 5 LCD display 6 Front panel digital instruments 6 Connection panel 7 Lid open closure 7 Preparations for use all instruments 8 Batteries 8...

Page 4: ...1 to 20 n n Audible buzzer disable n n Lead null to 9 0 0 5 for MIT310A n n 0 5 n n Stored test lead null n n n n n Resistance 10 to 1M resistance range n n INTRODUCTION GENERAL DESCRIPTION 4 Thank yo...

Page 5: ...Certificate of test 1 x Calibration certificate 1 x CD containing user manual 1 x Printed quick start guide 1 x USB lead MIT330 only 1 x SP4F switched probe MIT320 and MIT330 only CASE CONTENTS 5 MIT...

Page 6: ...st lock Measurment ranges Analogue pointer Battery condition indicator Warning refer to user guide Multi fuction analogue range display Fuse blown indicator Continuity indicator Test lock warning Infi...

Page 7: ...panel tab 1 2 Fold away underneath instrument 2 3 and push into retaining slot 4 Connection Panel MIT300 Series test lead connections An interlock sliding panel prevents connection of unsafe test lea...

Page 8: ...nge all the time the TEST button is locked down Voltage warning MIT300 only If a voltage exists on the circuit under test the instrument will emit a warning beep If a voltage greater than 50 V exists...

Page 9: ...operator should refer to the user manual for further information LRange lock Displayed at any time the TEST button is locked in the on position zLead nulled Indicates that the test lead resistance ha...

Page 10: ...istance should be zeroed out This ensures any measurement does not include the resistance of the test leads 1 Connect the test leads to the instrument and firmly short both leads together 2 Wait for t...

Page 11: ...st check which looks for a resistance greater than 1 0 M and will buzz when found If a voltage is detected the continuity test will stop the instrument will not be damaged The continuity buzzer will t...

Page 12: ...test probes to the isolated circuit under test 3 Press and hold the TEST button allow the display to settle The reading will remain displayed for a few seconds after the push button is released 4 Rel...

Page 13: ...o OFF and the threshold value set to 1 00 M To adjust 1 Switch the instrument to SETUP 2 Repeatedly press the TEST button until the display shows RES 3 The initial insulation alarm level is set to 1 0...

Page 14: ...ion of the needle over the taper symbol will indicate battery condition Highest charge to the right Note NiMH or NiCAD rechargeable batteries show a lower charge than alkaline batteries and my not giv...

Page 15: ...ance Continuity test MIT310A Continuity range extends from 0 00 to 2 k Ranges are not auto ranging Manual selection of each range is necessary Test currents are 200 mA to 2 20 mA to 20 2 mA to 200 200...

Page 16: ...1 P2 Phase number each test can be stored under a particular phase P1 P2 or P3 Unique test number each test result is assigned a unique test number from 0 to 1999 This is automatic and cannot be adjus...

Page 17: ...ar graph shows the progress of the deletion 17 2 Select job reference number using keys then press NEXT Hold the key down to scroll quickly through the numbers 3 Select distribution board number b01 0...

Page 18: ...ay not give much warning before becoming exhausted To replace batteries Warning Do not switch the instrument on with the battery cover removed 1 Switch off the instrument and disconnect the instrument...

Page 19: ...he fuse symbol findicates a ruptured fuse The fuse is located behind the rear cover The rear cover must not be opened if test leads are connected To avoid the possibility of shock disconnect the batte...

Page 20: ...d IEC 364 Accuracy at 20 C MIT300 310 320 330 3 2 digits up to 10 M 5 up to 100 M 30 up to full scale MIT310A 2 5 of scale length or 30 of reading 200 k to 10 M Terminal characteristics Continuity ran...

Page 21: ...perates as voltmeter when test button not pressed Test inhibit If more than 50 volts is detected testing will be inhibited Autopower down Auto power down operates after 10 minutes if left in standby m...

Page 22: ...E M C In accordance with IEC61326 1 Operational uncertainties Refer to www megger com Order Code Included Accessories Test lead set red black with crocodile clips 1002 001 USB download lead MIT330 on...

Page 23: ...eturning your product to Megger UK and USA service centres 1 When an instrument requires recalibration or in the event of a repair being necessary a Returns Authorisation RA number must first be obtai...

Page 24: ...nce with Directive 2014 53 EU Other equipment manufactured by Megger Instruments Limited described in this user guide is in compliance with Directives 2014 30 EU and 2014 35 EU where they apply The fu...

Page 25: ...megger com Megger Pty Limited Unit 1 11 21 Underwood Road Homebush NSW 2140 T 61 0 2 9397 5900 F 61 0 2 9397 5911 E ausales megger com Megger Limited Unit 106 550 Alden Road Markham Ontario L3R6A8 Ca...

Reviews: