Setup options
Setup options
The Setup options allow the MFT to be configured to best suit the type of testing to which it will be used.
To enter Setup, right (secondary) range knob to
Setup. Set the left (primary) range knob to any function other than OFF.
The display will show
VER
and the software version number. It will then change to the first message in the list below:
Message Function
Options
Factory setting
RST
Restore factory settings
NO / YES
NO
INS*
1
Insulation limit alarm –
Buzzer sounds if result is
higher than limit set
0.5, 1, 2, 3, 4, 5,7, 10, 50,
100, 500 MΩ
1 MΩ
LOC
Insulation test lock.
ON / OFF
ON
bUZ
Continuity limit alarm –
Buzzer sounds if result is
below than limit set
0.5, 1, 2, 5, 10, 50, 100 Ω
2 Ω
ISC*
2
Continuity test current
15 mA / 200 mA
200 mA
REV
Auto reverse continuity test
ON / OFF
OFF
looP
Loop test lead compensation 0 – 0.3 ohms
0.07 Ω
Not used on
this model.
LAS
Loop test AUTO start
ON/OFF
OFF
L-PE 2Hi
Enable/Disable high current
loop test
ON/OFF
ON
L-PE 2Lo
ON/OFF
ON
RAS
GFCI AUTO start
ON/OFF
OFF
UL
Touch voltage limit
25 V / 50 V / 60 V
50 V
OFF
Auto switch OFF in minutes
2 m / 20 m
20 minutes
bAC
Backlight AUTO mode/
manual mode
In Auto mode, backlight
comes on when at start
of test, end of test, range
change etc.
Auto / nor
Auto
bAt
Alkaline or NiMH selection
1.5 V or 1.2 V
Depending on instrument
StR
Store mode
IN = Internal
Bt = Bluetooth
®
only
IN / bT
IN
bt
Bluetooth
®
pairing
bt1, bt2, bt3, bt4, bt5
bt1
< >
Searching for pair
To scroll down the options, press the
button. Each option will be is played in sequence.
To change the setting of each function, for example, INS limit alarm from 1 MΩ to 2 MΩ, use the right hand TEST
and LOCK keys (also marked with UP/DOWN arrows).
Changing an option will set the LOCK symbol and warning triangle flashing.
To save the change press the left LOCK button
To exit SETUP, turn the right range knob away from
All settings can be restored to the factory defaults by setting RST to YES. Saving this setting will reset all options to
default.
The RST will then set back to NO.