Earth Test Ranges:
Intrinsic accuracy:
± 2.0% ± 3 digits.
ART Method
± 5.0% ± 3 digits.
Stake-less Method
± 7.0% ± 3 digits.
Resolution: 0.01
W
EN61557 Range:
1.0
W
to 1.99 k
W
Current:
0.45 mA or 4.5 mA.
Noise Rejection:
20 V pk/pk (7V rms).
Max Probe Resistances:
Rp, Rc = 100 k
W
at 50 Volts.
Max Service error:
± 20% ± 3 digits
Current (via Clamp meter):
Intrinsic accuracy:
± 5.0% ± 3 digits.
Resolution:
0.1 mA
EN61557 Range:
0.5 mA – 199 A.
Max Service error:
± 10% ± 2 digits.
Power consumption:
Nominal Minimum:
60 mA (Voltage range with no input voltage)
Nominal maximum:
350 mA (Active Insulation test set to 1000 V / 1 M
W
)
Range:
-20 °C to +55 °C
Temperature (via 3rd Party Module):
Intrinsic accuracy:
± 1.0% ± 2 digits.
Resolution:
1 °C
Range:
-20 °C to +100 °C
Environmental Specification.
Temperature
Operational range:
-10 °C to +55 °C
Storage range:
-25 °C to +70 °C
Humidity
Operating Humidity:
90% R.H. at +40 °C max.
Altitude
2000m to full safety specification.
Weight:
1000g ±10% including batteries but excluding test leads, accessories and carry case.
IP rating:
IP54
Power Supply.
Battery: Primary
6 x 1,5 V cells IEC LR6 type (AA alkaline,).
Battery: Rechargeable:
6 x 1.2 VNiMH cells EC HR6.
Display shows Rechargeable [NiMH] when the battery type is changed in setup option (Section 10).
1) Safety / EMC.
EMC In accordance with IEC61326 edition 2.
Locations:
Class B locations.
Safety in accordance with BS EN 61010 -1 2010 + 61010 -30:2010
Installation Category: 600 V Cat III / 300 V Cat IV. (Max Phase to Phase 550 V)
In addition Switch probe and test leads are designed to meet IEC 1010-031:2008 , Double insulated to Installation Category III , 300V
phase to earth , 500V phase to phase.