5.2 Auto stop
1. Turn rotary switch to select the inductive mode position.
5.2.1 First test
Inductive test screen appears, ‘Inductive Mode’ shows in
secondary field for a short time then disappears.
Note:
Soft key 1 toggles MANUAL / AUTO.
2. Select AUTO on soft key 1.
3. Connect C1-C2 and P1-P2 to the instrument and the unit
under test.
C and P indicators show red background with
for no
continuity or green background with
for good continuity.
Proceed when both show
. The continuity indicators are
active during the test and will update if continuity is lost.
Tests, Inductive mode
www.megger.com
DLRO2
18