www.megger.com
DLRO10HD and DLRO10HDX
17
Tests
Continuous Test
AUTO
3s
2.5
Ω
250
mΩ
µA
100
2.5
kΩ
1mA
10mA
100
mA
1A
1A
10A
10A
250
Ω
25
Ω
2.5
Ω
250
mΩ
25
mΩ
2.5
mΩ
Tip:
Use Clamp test leads (optional accessory).
1. Press
TEST
.
Test starts (LED shows red).
Arrow direction animated (based on current direction)
and plateau.
TEST
2. Continuous measurements are taken on the connected
test piece.
Tests are made every three seconds.
Last test record is shown while the next test is in
progress.
3.
DLROHD10X only
: Save test results (if required).
At any point press
(acknowledged by a beep
sound).
Test results are saved until the test is stopped or the
memory is full (2000 records).
4. Press
TEST
to stop the test.
DLROHD10X only:
If
was pressed in Step 3, Test
results are saved and the memory slot number is shown.
The Save screen is shown for three seconds.
5.
DLROHD10X only:
If
was not pressed in Step 3,
save test results if required.
Press
. The last complete test result is saved and its
memory slot number is shown.
The Save screen is shown for three seconds.