DLRO10HD and DLRO10HDX
10
Overview
Resolution and Accuracy
Test current accuracy ±10%
Voltmeter input impedance >200 kΩ
Maximum lead resistance at 10 A <100 mΩ
Test Current Resistance Range
Resolution
(as shown)
Basic Accuracy* Full Scale
Voltage
Max. Power
Output
100 µA
0 to 2.5 kΩ
0.1 Ω
±0.2% ±200 mΩ 25 mV
25 µW
100 µA
0 to 250 Ω
0.01 Ω
±0.2% ±20 mΩ
25 mV
2.5 µW
1 mA
0 to 25 Ω
1 mΩ
±0.2% ±2 mΩ
25 mV
25 µW
10 mA
0 to 2.5 Ω
0.1 mΩ
±0.2% ±200 µΩ 25 mV
250 µW
100 mA
0 to 250 mΩ
0.01 mΩ
±0.2% ±20 µΩ
25 mV
2.5 mW
1 A
0 to 25 mΩ
1 µΩ
±0.2% ±2 µΩ
25 mV
25 mW
10 A
0 to 2.5 mΩ
0.1 µΩ
±0.2% ±0.2 µΩ
25 mV
0.25 W
1 A**
0 to 2.5 Ω
0.1 mΩ
±0.2% ±200 µΩ 2.5 V
2.5 W
10 A**
0 to 250 mΩ
0.01 mΩ
±0.2% ±50 µΩ
2.5 V
25 W
* Basic accuracy stated assumes forward and reverse measurements.
** Higher 2.5 W (1 A) and 25 W (10 A) power outputs (
G
shows).
Inductive or Uni-directional mode can introduce an undefined error if an external EMF is present.
Basic accuracy at reference conditions.