Display:
Alpha numeric LCD (130 mm x 35 mm) giving test information and a large
(20 mm) 3
1
⁄
2
digitLCD, maximum reading 1999
Instrument Protection:
Meets the general requirements of IP54
Temperature Effect:
<±0,1%/°C over the temperature range -10°C to +40°C
Temperature Range:
Operating:
-10°C to +40 °C
Storage:
-20°C to +60°C
Humidity:
Operating:
90% RH max. at 40°C
Flash Test:
3 kV a.c.
Voltage Withstand:
In the event of a system fault the instrument will withstand 240 V a.c.
applied between any two terminals.
Compliance with Standards:
BS 7430 (1992)
BS7671 (1992) NFC 15-100
VDE 0413 Part 7 (1982)
IEC364
Power Supply:
(i) Internal rechargeable sealed lead acid cells 12 V nominal, 2,6 Ah
capacity. Battery voltage range over which basic accuracy is
maintained, 11,0 V to 13,5 V.
Battery life:
Typically 5 Hours continuous use
Battery charging time:
6 hours max. (from completely exhausted).
Charging supply required:
100 V to 130 V or 200 V to 260 V a.c. 50 Hz/60 Hz.
Power consumption:
25 VA