Battery type:
DET3TA, DET3TC, DET3TD, DET4TD2, DET4TC2
- 8 off AA (LR6) dry cells
DET4TCR2, DET4TR2
– 8 off AA (LR6) NiMH rechargeable cells
Operating temperature range:
-15 °C to +55 °C / 5 °F to 131 °F
Storage temperature range:
-40 °C to +70 °C / -40 °F to 158 °F
Safety
Complies with the requirements of IEC61010-1 100V CAT IV
between terminal pairs.
EMC
In accordance with IEC61326-1
Operational uncertainties
Refer to www.megger.com
Standards compliance
Complies with the requirements of KEMA K85B
Complies with the following parts of EN61557, Electrical safety
in low voltage distribution systems up to 1000 V a.c. and
1500 V d.c. – Equipment for testing, measuring or monitoring
of protective measures”.
Part 1 – General requirements
Part 5 – Resistance to earth
Dimensions:
203 mm x 148 mm x 78 mm / 8 in x 5.8 in x 3 in
Weight:
1 kg / 2.2 lb
MCC CLAMP SPECIFICATIONS
Measurement range:
1 mA to 1000 A Transformation ratio: 1000 : 1
Output signal:
1 mA / A (1 A @ 1000 A)
Accuracy
1mA to 100 mA : ≤3% of reading
0.1 A to 1 A : ≤2% of reading
1 A to 10 A : ≤1% of reading
10 A to 100 A : ≤0.5% of reading
Operating temperature
-20 °C to +50 °C / -4 °F to 122 °F
0% to 85% RH at +35 °C / 95 °F
Storage temperature
-20 °C to +70 °C / -4 °F to 158 °F
Jaw opening:
50 mm maximum (2.0 in)
Maximum conductor size:
50 mm (2.0 in)
Insulation:
Double insulation
Voltage rating:
CAT IV 600 V
Current rating:
1000 A for 20 minutes
Ingress protection:
IP40
Safety
IEC61010-2-032
EMC
IEC61326-1
Dimensions:
105 mm x 216 mm x 45 mm (4.1 in x 8.5 in x 1.8 in)
Weight:
0.5 kg approximately (1.2lb)