Setting up the DLRO2 and DLRO2X
www.megger.com
DLRO2 and DLRO2X
20
4.2.5 Setting test limits (DLRO2X)
Switch the rotary switch to the settings mode
(Refer
to 9.1 General settings on page 28).
1. Use the arrow keys to select the Lower Test Limit
and click the
TICK button (
)
to select.
2. Use the soft keys 1 or 4 to move the selector left
or right to choose the Units of measurement and
the limit required.
Use the up and down arrows (soft keys 2 and 3)
to select the numbers..
3. When the selection is complete use soft key 4
to click the TICK (
) and to save your test limit
settings.
4. Go back to the main Settings menu and do the
same for the Upper Test Limit.