After starting a test the user will see confirmation of successful continuity
testing on the LCD. If the dashes are not shown connecting 1 & 2 C and/or P
leads connection has not been detected.
C 1---2
P 1---2
The resistance result is displayed on the large 5-digit digit display in either Ω,
mΩ or µΩ ranging from 2500.0 Ω to 0.1 µΩ.
The result in bidirectional modes is the average of two readings shown by
the two secondary displays with arrows to indicate direction of current flow.
The large arrow at the top of the display between ‘C1’ and ‘C2’ indicates
measurement current flow.
LCD Icons
Battery icons with battery charge level indicator
Line power connected
Temperature extreme limit warning (can inhibit battery
charging if ambient temperature is too low/high)
Noise present above instrument noise threshold
Warning – refer to user guide
HV warning – used to indicate voltage present
Directional arrow to indicate current flow above C
indicator
Directional arrow to indicate current flow above P
indicator
Arrows used to show direction of current during test