![Medtronic MIDAS REX LEGEND GOLD Preventive Maintenance/Service Manual Download Page 9](http://html.mh-extra.com/html/medtronic/midas-rex-legend-gold/midas-rex-legend-gold_preventive-maintenance-service-manual_1763058009.webp)
7
4
th
Scheduled Maintenance
(See Section 2 for Details)
Date
Performed:_________________
By:_________________
Level:
A, 12 months
B, 16 months
C, 20 months
D, 24 months
Motor
Attachments: Straight, Footed,
Perforator, Jacobs Chuck,
Telescoping
Attachments:
Angled, Contra, Metal Cutting
Foot Control, Regulator
Requires factory level service.
Please call 817-788-6440 or toll free
1-800-335-9557 to schedule.
Sent out on ____________________
Returned on ___________________
Requires factory level service.
Please call 817-788-6440 or toll free
1-800-335-9557 to schedule.
Sent out on ____________________
Returned on ___________________
NOTE:
Telescoping tubes are multi-use
disposable. Discard if excessive heat
or vibration is present.
Requires factory level service.
Please call 817-788-6440 or toll free
1-800-335-9557 to schedule.
Sent out on ____________________
Returned on ___________________
Contra-Angled Attachment —
Replace
Please call 817-788-6400 or toll free
1-800-433-7080 to order.
Ordered on ____________________
Air
fi ttings and supply hose do not
show any signs of blockage, damage or
wear.
Pressurize the foot control or regulator
and confi rm gage functionality.
NOTE:
Th
e pressure gages should
be placed on your facilities routine
calibration schedule.
Pressurize the foot control and press
Pressure Relief button. Ensure that it
releases the compressed gas.
Confi rm that the Finger/Foot Control
Switch is functional.
(In
the position, the foot pedal
will be depressed.
In
the position, the foot pedal
will be released.)
Legend and Triton Port Plugs are intact
and without damage.
Foot pedal moves freely and there is no
build-up of debris.
Motor does not continue to run when
foot pedal is released.
If any damage, excessive rust/corrosion, or wear is observed with the product, or if action is required beyond cleaning,
DO NOT
use the product and return to MPSS for service.