Medtronic MIDAS REX LEGEND GOLD Preventive Maintenance/Service Manual Download Page 14

12

9

th

 Scheduled Maintenance 

(See Section 2 for Details)

 Date 

Performed:_________________ 

By:_________________

Level: 

  A, 27 months 

 B, 36 months 

 C, 45 months 

 D, 54 months

Motor

Attachments: Straight, Footed, 

Perforator, Jacobs Chuck, 

Telescoping

Attachments:

Angled, Contra, Metal Cutting

Foot Control, Regulator

  Exhaust hose does not have any nicks, 

cracks, holes, tears, or other visible 
signs of wear.

  Exhaust hose does not remain oily aft er 

wiping.

 Lubricant/Diff user connection does 

not show any signs of damage that 
would prevent the lubricant/diff user 
from fi tting properly.

  Air Fitting does not exhibit any signs of 

damage.

  Etching remains legible.

  Connect a straight or footed 

attachment to the motor and run for 
5 minutes. Motor does not become 
uncomfortably hot to the touch and 
there is no oil leaking from the motor 
collet.

  (Gold Touch only) Motor does not 

continue to run when the fi nger control 
lever is released.

  (Gold Touch only) Finger control lever 

and safety slide has no signs of debris 
or damage, and the lever moves freely. 

  Etching and color ring remain legible 

and distinguishable.

  Attachment tube and base are clean 

and free of debris.

  Attachment has no loose components.

  (Straight/Footed) Attachment tip 

shows no signs of wear or deformation.

  Footed attachment foot is not bent.

  Collet on Perforator or Jacobs Chuck 

Drive Attachment shows no signs of 
wear or deformation.

  Connect the attachment to the motor 

and run for 5 minutes. Motor case, 
attachment base and tip do not become 
uncomfortably hot to the touch.

  Repeat for each straight, footed, 

telescoping, perforator, and Jacobs 
Chuck drive attachment in this system. 

NOTE:

 Telescoping tubes are multi-use 

disposable. Discard if excessive heat 
or vibration is present.

  Etching and color ring remain legible 

and distinguishable.

  Attachment tube and base are clean 

and free of debris.

  Attachment has no loose components.

  Attachment tip shows no signs of wear 

or deformation.

  Etching remains legible.

  Connect the angled or metal cutting 

attachment to the motor and run for 
5 minutes. Motor case, attachment 
base, knuckle and tip do not become 
uncomfortably hot to the touch.

  Head to tube joint for the 

contra-angled attachment is stable, i.e., 
no play.

  Connect the contra-angled attachment 

to the motor and run for 1 minute. 
Motor case, contra-angled attachment 
base, knuckle and tip do not become 
uncomfortably hot to the touch.

  Repeat for each angled, contra-angled 

or metal cutting attachment in this 
system. 

 Air 

fi ttings and supply hose do not 

show any signs of blockage, damage or 
wear.

  Pressurize the foot control or regulator 

and confi rm gage functionality.  

NOTE

: Th

  e pressure gages should 

be placed on your facilities routine 
calibration schedule.

  Pressurize the foot control and press 

Pressure Relief button. Ensure that it 
releases the compressed gas.

 Confi rm that the Finger/Foot Control 

Switch is functional. 

 (In 

the   position, the foot pedal 

will be depressed. 

 In 

the   position, the foot pedal 

will be released.)

  Legend and Triton Port Plugs are intact 

and without damage. 

  Foot pedal moves freely and there is no 

build-up of debris.

  Motor does not continue to run when 

foot pedal is released. 

If any damage, excessive rust/corrosion, or wear is observed with the product, or if action is required beyond cleaning, 

DO NOT

 use the product and return to MPSS for service.

Summary of Contents for MIDAS REX LEGEND GOLD

Page 1: ...MIDAS REX LEGEND Pneumatic High Speed System Preventive Maintenance Service Manual ...

Page 2: ...d Maintenance 7 5th Scheduled Maintenance 8 6th Scheduled Maintenance 9 7th Scheduled Maintenance 10 8th Scheduled Maintenance 11 9th Scheduled Maintenance 12 10th Scheduled Maintenance 13 11th Scheduled Maintenance 14 12th Scheduled Maintenance 15 Section 2 Instructions for Completing Preventive Maintenance Steps 17 Scheduled Maintenance Intervals 1 2 3 5 6 7 9 10 11 18 Scheduled Maintenance Inte...

Page 3: ... upon how often it is used and the conditions to which it is subjected such as cleaning and sterilization The demands on pneumatic systems can vary greatly from facility to facility Twelve maintenance intervals have been defined The time interval is determined by tool usage per year per system and has been identified as levels A B C or D in accordance with the following chart Maintenance Level No ...

Page 4: ... in accordance with the Twelve Maintenance Intervals Your Midas Rex Legend Pneumatic System has a limited warranty for a period of 1 year from the date of purchase for Gold Gold Touch andPlatinumMotors Hoses PneumaticControlUnitwithPressureHose Regulator Adapters andInstrumentCasesand Trays The Attachments and Telescoping Tubes have a limited warranty for a period of 90 days from the date of purch...

Page 5: ...ection 1 Preventive Maintenance Service Schedule 12 Maintenance Intervals Customer Performed Maintenance MPSS Performed Maintenance Date of System Purchase __________________ Maintenance Level ___________ ...

Page 6: ...ping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the a...

Page 7: ...oping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the ...

Page 8: ...oping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the ...

Page 9: ...____________________ Returned on ___________________ Contra Angled Attachment Replace Please call 817 788 6400 or toll free 1 800 433 7080 to order Ordered on ____________________ Air fittings and supply hose do not show any signs of blockage damage or wear Pressurize the foot control or regulator and confirm gage functionality NOTE The pressure gages should be placed on your facilities routine ca...

Page 10: ...coping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the...

Page 11: ...coping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the...

Page 12: ...coping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remains legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect th...

Page 13: ...__________________ Returned on ___________________ NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Requires factory level service Please call 817 788 6440 or toll free 1 800 335 9557 to schedule Sent out on ____________________ Returned on ___________________ Contra Angled Attachment Replace Please call 817 788 6400 or toll free 1 800 433 7080 to o...

Page 14: ...coping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the...

Page 15: ...scoping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect th...

Page 16: ...scoping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect th...

Page 17: ... ____________________ Returned on ___________________ Contra Angled Attachment Replace Please call 817 788 6400 or toll free 1 800 433 7080 to order Ordered on ____________________ Air fittings and supply hose do not show any signs of blockage damage or wear Pressurize the foot control or regulator and confirm gage functionality NOTE The pressure gages should be placed on your facilities routine c...

Page 18: ...16 ...

Page 19: ...Section 2 Instructions for Completing Preventive Maintenance Steps 17 ...

Page 20: ...ct the etching and ensure that it remains legible 6 Heat and Seal Check Attach and lock a straight or footed attachment onto the motor Check to ensure that there is a tactile feel and an audible click when the attachment is locked on the motor Connect the motor to a Foot Control unit attached to a compressed gas source Run the motor for 5 minutes Feel the motor case and the attachment base and ens...

Page 21: ...n round with no evidence of deformation 5 Footed Visually inspect the foot The foot and or leg of the attachment should not be bent or damaged 6 Perforator and Jacobs Chuck Visually inspect the perforator collet for any signs of wear or deformation 7 Heat Check For telescoping attachment insert an appropriate telescoping tube into the base Attach and lock the attachment onto the motor Check to ens...

Page 22: ...nect the motor to a Foot Control unit attached to a compressed gas source Run the motor for 5 minutes Feel the motor case the attachment base the knuckle angled area for angled attachment only and the attachment tip to ensure they are not uncomfortably hot to the touch 6 Contra Visually inspect the head to tube joint for the contra angled attachment for stability i e there is no loosening of this ...

Page 23: ...libration date 3 Foot Control Pressure Relief Button Check Connect the foot control to a compressed gas source Pressurize the foot control Turn off the compressed gas source Press the Pressure Relief Button Confirm that the compressed gas has been released and the gage on the foot control drops to zero 4 Foot Control Finger Foot Control Switch Functionality Check Place the switch in the position C...

Page 24: ...Check Attach a Gold or Platinum motor to the foot control Connect the foot control to a compressed gas source Depress the foot control and run the motor for a short duration Release the foot control and observe the motor NOTE If the motor continues to run after the foot pedal control lever is released DO NOT use and return the foot control to MPSS for service ...

Page 25: ...tion Telescoping tubes are multi use disposable and do not receive scheduled maintenance Contra Angled Attachment The Contra angled attachment does not contain user or factory serviceable components and must be replaced at the 4th 8th and 12th maintenance intervals Please call 817 788 6400 or toll free 1 800 433 7080 to order Foot Control and Regulator 1 Visually inspect the air fittings and suppl...

Page 26: ...is released 5 Foot Control Visually inspect the Legend and Triton port plugs for any signs of wear or damage and confirm that they remain connected to the foot control 6 Foot Control Visually inspect the foot control pedal for any build up of debris or indication of damage Ensure the pedal moves freely 7 Foot Control Pedal Check Attach a Gold or Platinum motor to the foot control Connect the foot ...

Page 27: ...ts associated with any of the factory level service requirements are not included in the limited warranty Extended Maintenance Agreements are available Please refer to the first page insert for the Business Reply Card to obtain further information Telescoping tubes are multi use disposable and do not receive scheduled maintenance Contra Angled Attachment The Contra angled attachment does not conta...

Page 28: ...26 Notes ...

Page 29: ...gical Solutions Telephone 817 788 6400 Toll free 1 800 433 7080 Fax 817 788 6489 Repairs Medtronic Powered Surgical Solutions Telephone 817 788 6440 Toll free 1 800 335 9557 Fax 817 788 6441 EC Authorized Representative Medtronic B V Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Patent Pending Pana is a trademark of NSK America Corporation 2014 Medtronic Inc All rights reserved Printed in U...

Reviews: