12
9
th
Scheduled Maintenance
(See Section 2 for Details)
Date
Performed:_________________
By:_________________
Level:
A, 27 months
B, 36 months
C, 45 months
D, 54 months
Motor
Attachments: Straight, Footed,
Perforator, Jacobs Chuck,
Telescoping
Attachments:
Angled, Contra, Metal Cutting
Foot Control, Regulator
Exhaust hose does not have any nicks,
cracks, holes, tears, or other visible
signs of wear.
Exhaust hose does not remain oily aft er
wiping.
Lubricant/Diff user connection does
not show any signs of damage that
would prevent the lubricant/diff user
from fi tting properly.
Air Fitting does not exhibit any signs of
damage.
Etching remains legible.
Connect a straight or footed
attachment to the motor and run for
5 minutes. Motor does not become
uncomfortably hot to the touch and
there is no oil leaking from the motor
collet.
(Gold Touch only) Motor does not
continue to run when the fi nger control
lever is released.
(Gold Touch only) Finger control lever
and safety slide has no signs of debris
or damage, and the lever moves freely.
Etching and color ring remain legible
and distinguishable.
Attachment tube and base are clean
and free of debris.
Attachment has no loose components.
(Straight/Footed) Attachment tip
shows no signs of wear or deformation.
Footed attachment foot is not bent.
Collet on Perforator or Jacobs Chuck
Drive Attachment shows no signs of
wear or deformation.
Connect the attachment to the motor
and run for 5 minutes. Motor case,
attachment base and tip do not become
uncomfortably hot to the touch.
Repeat for each straight, footed,
telescoping, perforator, and Jacobs
Chuck drive attachment in this system.
NOTE:
Telescoping tubes are multi-use
disposable. Discard if excessive heat
or vibration is present.
Etching and color ring remain legible
and distinguishable.
Attachment tube and base are clean
and free of debris.
Attachment has no loose components.
Attachment tip shows no signs of wear
or deformation.
Etching remains legible.
Connect the angled or metal cutting
attachment to the motor and run for
5 minutes. Motor case, attachment
base, knuckle and tip do not become
uncomfortably hot to the touch.
Head to tube joint for the
contra-angled attachment is stable, i.e.,
no play.
Connect the contra-angled attachment
to the motor and run for 1 minute.
Motor case, contra-angled attachment
base, knuckle and tip do not become
uncomfortably hot to the touch.
Repeat for each angled, contra-angled
or metal cutting attachment in this
system.
Air
fi ttings and supply hose do not
show any signs of blockage, damage or
wear.
Pressurize the foot control or regulator
and confi rm gage functionality.
NOTE
: Th
e pressure gages should
be placed on your facilities routine
calibration schedule.
Pressurize the foot control and press
Pressure Relief button. Ensure that it
releases the compressed gas.
Confi rm that the Finger/Foot Control
Switch is functional.
(In
the position, the foot pedal
will be depressed.
In
the position, the foot pedal
will be released.)
Legend and Triton Port Plugs are intact
and without damage.
Foot pedal moves freely and there is no
build-up of debris.
Motor does not continue to run when
foot pedal is released.
If any damage, excessive rust/corrosion, or wear is observed with the product, or if action is required beyond cleaning,
DO NOT
use the product and return to MPSS for service.