19
Attachments: Straight, Footed, Perforator, Jacobs Chuck, and Telescoping
Telescoping tubes are multi-use disposable and do not receive scheduled maintenance.
1. Visually inspect the etching and color ring to ensure it remains legible and the color is distinguishable.
2. Visually inspect the attachment tube and base for any signs of debris or damage.
3. Visually and physically inspect (pulling and twisting by hand) the attachment for any loose or loosening components. If
any components are loose,
DO NOT
use and return the attachment to MPSS for service.
4. (Straight/Footed) Visually inspect the attachment tip for any signs of wear. Th
e tip should remain round with no evidence
of deformation.
5. (Footed) Visually inspect the foot. Th
e foot and/or leg of the attachment should not be bent or damaged.
6. (Perforator and Jacobs Chuck) Visually inspect the perforator collet for any signs of wear or deformation.
7. Heat
Check
•
For telescoping attachment, insert an appropriate telescoping tube into the base.
•
Attach and lock the attachment onto the motor. Check to ensure that there is a tactile feel and an audible click when
the attachment is locked on the motor.
•
Connect the motor to a Foot Control unit attached to a compressed gas source.
•
Run the motor for 5 minutes.
•
Feel the motor case, the attachment base, and the attachment tip (except Telescoping) to ensure they are not
uncomfortably hot to the touch.
8. Repeat steps 1-7 for each straight/footed/perforator/Jacobs Chuck attachment that is part of this system.