INSTRUCTIONS FOR USE:
To charge the clipper body:
Your surgical clipper may initially arrive with a limited charge. Before use,
it should be charged for 12 hours. To recharge, place clipper in charging
adapter. Make sure the indicator light is on to ensure proper connection to
the charging adapter. A completely depleted battery will require 12 hours
of charging time. Damage will not result from over-charging the battery.
To attach blade:
Make sure that the on/off switch is in the off position.
Line up the small line on the side of the clipper blade with the first
(upward) small line on the clipper handle. Fit the blade on the handle by
pushing downward until the line on the clipper blade lines up with the
second (lower) line on the clipper and it will click into place.
To use clipper:
Patient’s skin should be clean. Attach blade per instructions. Place On/Off
switch in the on position. Hold the clipper body so that the blade assembly
is flat against the patient’s skin and use short gentle strokes. For best
results, clip against the direction of hair growth. For problem areas, stretch
the skin taut while clipping. When finished, place On/Off switch in the off
position. Remove blade when not in use.
WARNINGS:
When using this surgical clipper, basic precautions should always be
followed. Please read and understand all safety information before using.
1.
Use this clipper only for its intended use as explained in this insert.
2.
The clipper must only be used with the charger provided.
3.
Do not submerge charging adapter in water or any other liquid. Do not
store the charging adapter near water. Should the charging adapter
fall into water or any other liquid, immediately unplug the charger
from the wall unit before attempting to retrieve the adapter.
4.
If hands are wet, do not attempt to plug or unplug the charger from
an electrical unit.
5.
Keep the charging adapter cord away from heated surfaces.
6.
Do not replace the supply cord.
7.
Always unplug the charger prior to cleaning.
8.
If blade becomes damaged, do not use. Replace blade according to
instructions.
9.
Do not sterilize blade or charger.
10.
To avoid risk of fire or explosion, do not use near flammable
anesthetics, aerosol sprays, or oxygen-administering equipment other
than those administered via nasal cannulas or masks.
11.
This product has been tested according to Electromagnetic
Compatibility Testing per IEC60601-1-2. This unit does not run the risk
of discharging radiation and affecting other equipment. If undesirable
performance is experienced, discontinue use of product.
12.
Do not use or charge the clipper if it has a damaged plug or cord.
13.
Do not charge in the vicinity of the patient.
14.
To turn off charging adapter, disconnect the plug from the main
power source.
15.
WARNING:
No modification of this equipment is allowed.