Approximate dimensions of the water filter are shown in figure 2.

•  For wall mounting: 

Mount the filter assembly on the wall where the pressure gauges can be read easily. 

Allow a minimum of 30 inches under the assembly for removal of the filter housings and 2.5 inches on 

each side for connection of hoses.

•  For cart mounting:

 Mount the filter assembly in the supplied cart. The pressure gauges should be 

visible when the right hand access panel is removed.

•  Tap into the facility cold water supply line and install a 3/4 inch male hose fitting (not included).  This 

will require a qualified plumbing installer.

•  Using a customer supplied hose which is connected to the cold water supply, connect the other end 

of the hose to the male quick connect and attach the male quick connect to the female quick connect 

on the water filter assembly. This quick connected arrangement allows the use of the transfer pump for 

performing the water line disinfection.

•  Attach the 3/4 inch hose connection to one end of the supplied stainless steel hose and the other end 

to the water inlet of the reprocessor.

NOTE: Any time the final filter (0.2 micron) is changed, the water line must be  

           disinfected (Section 6)

CAUTION:  When changing filters, hand tighten the filter housing. Only use the filter  

                   wrench to remove housing.

NOTE: The pre-filter should be changed every three months, even if water pressure is good.  

            The final filter should be changed at a minimum every six months, even if the water  

            pressure is good.

 

  FILTER REPLACEMENT

The accumulation of particles in a filter can cause the water pressure to drop below the minimum required 
level (40 PSI). This depends on water quality which cannot be guaranteed by MEDIVATORS. The pressure 
drop across the filters is used to determine when the filter(s) should be changed. For example: A pressure 
differential of 5 PSI or greater between gauges on each side of a filter indicates a need for replacement. For 
example, if the pressure on gauge #1 is 35 PSI when the pressure on gauge #2 is 40 PSI, a filter change is 
required.
•  Turn off the water supply to the filtration system.
•  Press START on the reprocessor to remove the pressure from the filter housings.
•  Press STOP.
•  Remove filters, clean the housings with a mild general-purpose detergent and water, then rinse 

thoroughly.

•  Place the new 1-micron filter cartridge into the pre-filter housing. Both ends of this filter are open. 

Ensure that the large O-ring is seated and centered in the end of the filter when it is inserted into the 
housing. Thread the housing onto its header and hand tighten.

•  Place the new 0.2-micron bacterial-retentive filter cartridge into the final-filter housing. Only one end 

of this filter is open and has an O-rings. Ensure that the O-ring fits tightly on the peg in the top of the 
housing. Thread the housing onto its header and hand tighten.

5

3

Summary of Contents for CER

Page 1: ...5 to 3 GPM 9 4 LPM to 11 4 LPM filter system The cold water supply should not exceed 110 F 43 C FILTRATION SYSTEM NOTE Endoscope manufacturers recommended that endoscopes not be exposed to temperatures above 130 F 55 C NOTE A water shut off valve is required before the filter assembly It should be accessible to the users NOTE THE WATER SUPPLY IS REQUIRED TO BE TURNED OFF WHEN NOT IN USE 1 2 3 Gaug...

Page 2: ...r reorder CF01 0018 Filter Wrench CW03 0000 Stainless Steel Hose CH06 0000 Male quick connect fitting MC10 0042 Female Quick Connect MC10 0043 NOTE Do not remove the Male Female quick connectors in Figure 1 These are required to disinfect the filters and water line INSTALLATION 4 15 inches 16 inches Depth of unit is approximately 6 inches Figure 2 Water Filtration System Dimensions 2 ...

Page 3: ...ould be changed every three months even if water pressure is good The final filter should be changed at a minimum every six months even if the water pressure is good FILTER REPLACEMENT The accumulation of particles in a filter can cause the water pressure to drop below the minimum required level 40 PSI This depends on water quality which cannot be guaranteed by MEDIVATORS The pressure drop across ...

Page 4: ...ansfer pump into a one gallon container of high level disinfectant Press START on the reprocessor Turn on the transfer pump until the disinfectant solution starts flowing continuously in the basin Press STOP on the reprocessor Turn off the transfer pump Allow the reprocessor and the filters to soak for the time recommended on the product label of the high level disinfectant for high level disinfec...

Page 5: ...ctant and replace on header see Filtration Replacement Allow the filters to soak for the time recommended on the product label of the high level disinfectant for high level disinfection Turn on water supply to the filtration system Press START and allow the wash cycle to run in order to rinse the disinfectant out of the system When all of the water has drained press STOP 5 ...

Page 6: ...29 FAX 1 866 421 7696 www medivators com Manufactured in the USA by Medivators Inc 14605 28th Avenue North Minneapolis MN 55447 USA Toll Free 1 800 444 4729 Medivators BV Sourethweg 11 6422PC Heerlen The Netherlands Tel 31 45 5 471 471 Cantel Medical Asia Pacific Pte Ltd 1A International Business Park 05 01 Singapore 609933 Tel 65 6227 9698 Cantel Medical Devices China Co Ltd Unit 804 805 Innov To...

Reviews: