The Demand Valve is designed to have a product life span of 10 years, excluding abuse and/or 

damage to the device. 

 

10.   SPECIFICATION 

 

Inspiratory Resistance 

 

 

<1.5 kPa @ 200 l/min 

 

 

<0.25 kPa @ 10 l/min 

Expiratory Resistance 

 

 

<0.1kPa @ 12 l/min 

 

 

<0.6kPa @ 120 l/min 

Transport and Storage Temperature:  

-20°C to 60°C  

Operating Temperature:  

 

 

5°C to 40°C  

Supply Pressure @ 100l/min 

 

Nominal 400 kPa 

 

 

Maximum 600 kPa  

 

 

Minimum 280 kPa 

 

11. SPARE PARTS 

 

Description                                   

Part Number           

 

Inhalation Diaphragm                       

033-1031-00                           

 

 

12. RELEVANT APPLICABLE STANDARDS 

 

B.N.O.S. Meditech Ltd is an ISO 13485:2016 registered company. 

 

BS EN ISO 5356-1:2015 Anaesthetic and respiratory equipment, cones and connectors, cones and 

sockets. 

 

BS EN ISO 15001:2011 Anaesthetic and respiratory equipment. Compatibility with Oxygen. 

 

BS EN ISO 5359:2014+A1:2017 Low pressure hose assemblies for use with medical gases.  

 

BS 5682:2015 Probes (quick connectors) for use with medical gas pipeline systems. 

 

BS EN ISO 14971:2019 Medical Devices. Application of risk management to medical devices. 

 

BS EN ISO 15223-1:2016 Medical Devices. Symbols to be used with medical device labels, 

labelling and information to be supplied. General requirements.  

 

 

 

IMPORTANT NOTICE 

Manufacturer’s Warranty is for a period of 5 years and includes parts and labour. It does not 

include transport costs. The responsibility and cost of returning and collecting the unit from the 

manufacturer or their authorised representative is the owners.  

 

B.N.O.S Meditech Ltd reserves the right to change design without prior notice. 

 

Any disassembly of the device beyond that detailed in this manual will invalidate the 

warranty and the manufacturers disclaim any liability for products that have undergone 

unauthorised repair. 

 

 

 

Summary of Contents for OXYGEN DEMAND VALVE

Page 1: ...NTENDED USE OF DEVICE 2 3 TECHNICAL DESCRIPTION 2 4 WARNINGS 3 5 INSTRUCTIONS FOR USE 4 6 CLEANING AND DISINFECTION 5 7 MAINTENANCE 5 8 SERIAL NUMBER 6 9 PRODUCT LIFE SPAN 6 10 SPECIFICATION 6 11 SPAR...

Page 2: ...pieces require a 22mm female connection complying with BS EN ISO 5356 1 2015 2 TECHNICAL DESCRIPTION The principle of operation of the demand valve is similar to the original design of equipment used...

Page 3: ...8 ISSUE H 18 03 2021 The demand valve has been specifically designed to be used with low cost single use generic mouth piece and filter systems that are widely available using standard connections th...

Page 4: ...imes 4 9 The demand valve hose is fitted with medical gas specific connectors Do not interfere or modify the connector Particular attention must be paid during any form of maintenance to ensure that t...

Page 5: ...eleased 5 5 It is strongly advised to use a single use filter with the Demand Valve to minimise the risk of contamination of the device and a subsequent cross contamination between users Connect a uni...

Page 6: ...ice must be tested If required B N O S Meditech Ltd can advise on suitable test equipment 7 3 For Demand Valves in frequent use for example an Ambulance Service or a Maternity Unit personnel trained t...

Page 7: ...ment Compatibility with Oxygen BS EN ISO 5359 2014 A1 2017 Low pressure hose assemblies for use with medical gases BS 5682 2015 Probes quick connectors for use with medical gas pipeline systems BS EN...

Page 8: ...actured by B N O S Meditech Ltd 9 Fifth Avenue Bluebridge Ind Est Halstead Essex CO9 2SZ England Tel 44 0 1787 479475 Fax 44 0 1787 477747 E mail sales meditech uk com www meditech uk com EC Represent...

Reviews: