5.  INSTRUCTIONS FOR USE 

 

 

 

 

 

 

 

 

4. 

WARNINGS!  

4.1 Please read these instructions before use. Do not use the demand valve if you do 

not understand the instructions detailed in this user guide. 

4.2 

Do not use any form of grease or oil (hydrocarbon based substances) with 

this device. 

4.3 

Do not use this device near any source of ignition e.g. naked flame, electrically 

powered heaters, cigarettes etc.  

4.4 

 Do not smoke any products including tobacco when using this device. 

4.5 Medical Oxygen is a medicinal gas and its use must be supervised by correctly 

trained medical staff and/or prescribed for use where applicable by a suitably 

authorised medically trained individual e.g. Doctor. 

4.6 Do not use the “Press to Flush/Test” button when the device is in use. 
4.7 Ensure sufficient Oxygen is available for use before using the product.  
4.8 If using medical gas cylinders, ensure that they are adequately secured or stowed 

at all times. 

4.9 The demand valve hose is fitted with medical gas specific connectors. Do not 

interfere or modify the connector. Particular attention must be paid during any form 

of maintenance to ensure that the correct medical gas specific connector is fitted to 

the device. 

4.10 If used with a medical gas regulator, always ensure that the regulator cylinder valve 

is opened slowly. 

4.11 The device must not be used with harness or retained face mask. 
4.12 When using a regulator/cylinder, make sure that the regulator is capable of 

providing the Demand Valve with a flow of at least 100 l/min @ 4 bar of gas, to 

ensure that the device functions as intended. If in any doubt as to the suitability of 

the medical gas regulator to be used with the device please contact the 

manufacturer. 

4.13 When the demand valve is not in use, turn off the regulator where appropriate and 

disconnect the valve from the medical gas supply.  

4.14 Check the demand valve and supply hose regularly for leaks. Remove any leaking 

device from service immediately. 

4.15 Ensure that the hose is always arranged in such a way that it cannot be damaged 

and does not cause any form of obstruction or hazard. 

 

Summary of Contents for OXYGEN DEMAND VALVE

Page 1: ...NTENDED USE OF DEVICE 2 3 TECHNICAL DESCRIPTION 2 4 WARNINGS 3 5 INSTRUCTIONS FOR USE 4 6 CLEANING AND DISINFECTION 5 7 MAINTENANCE 5 8 SERIAL NUMBER 6 9 PRODUCT LIFE SPAN 6 10 SPECIFICATION 6 11 SPAR...

Page 2: ...pieces require a 22mm female connection complying with BS EN ISO 5356 1 2015 2 TECHNICAL DESCRIPTION The principle of operation of the demand valve is similar to the original design of equipment used...

Page 3: ...8 ISSUE H 18 03 2021 The demand valve has been specifically designed to be used with low cost single use generic mouth piece and filter systems that are widely available using standard connections th...

Page 4: ...imes 4 9 The demand valve hose is fitted with medical gas specific connectors Do not interfere or modify the connector Particular attention must be paid during any form of maintenance to ensure that t...

Page 5: ...eleased 5 5 It is strongly advised to use a single use filter with the Demand Valve to minimise the risk of contamination of the device and a subsequent cross contamination between users Connect a uni...

Page 6: ...ice must be tested If required B N O S Meditech Ltd can advise on suitable test equipment 7 3 For Demand Valves in frequent use for example an Ambulance Service or a Maternity Unit personnel trained t...

Page 7: ...ment Compatibility with Oxygen BS EN ISO 5359 2014 A1 2017 Low pressure hose assemblies for use with medical gases BS 5682 2015 Probes quick connectors for use with medical gas pipeline systems BS EN...

Page 8: ...actured by B N O S Meditech Ltd 9 Fifth Avenue Bluebridge Ind Est Halstead Essex CO9 2SZ England Tel 44 0 1787 479475 Fax 44 0 1787 477747 E mail sales meditech uk com www meditech uk com EC Represent...

Reviews: