background image

9

 

HOW YOUR SYSTEM WORKS 

 

Installing or Replacing the Battery 

Your 

MediSure

  Blood  Glucose 

Monitoring  System  is  packed  with 
one  3-volt  lithium  coin  battery 
that  needs  to  be  installed  before 
testing.    The  battery  will  last  for 
approximately  1,000  tests.  Always 
keep  spare  batteries  on  hand. 
When  the  battery  symbol  appears  on  the  display,  it 
means your battery is low. Although the meter will still 
provide accurate results, the battery should be replaced 
as soon as possible. After the battery is replaced, reset 
the date, and time.  
 
1.

 

Open the battery cover by pressing 

on it and sliding it in the direction 
of the arrows on the battery cover. 

2.

 

If replacing battery, remove the old 

battery. 

3.

 

Insert a new battery firmly with the 

positive (

+

) side up. The LCD will 

display “Set.”

 

4.

 

Reset the date, time. 

Summary of Contents for Blood glucose monitoring system

Page 1: ...ip 7 HOW YOUR SYSTEM WORKS 9 Installing or Replacing the Battery 9 Powering the Meter On and Off 10 Setting the Date Time 11 Alarm Settings 14 Reviewing Results in Memory 15 Viewing Date Time or Tempe...

Page 2: ...medical devices that require cleaning and disinfection 28 Information of CaviWipe disinfectant 29 IMPORTANT NOTES 30 Caring For Your Meter 30 Handling Storage 30 Explanation of Displays 31 ERROR MESS...

Page 3: ...o 960 test results with respective dates and times which allows you to monitor your Diabetes program efficiently Before testing please read this User s Guide thoroughly It contains all the information...

Page 4: ...TIONS FOR USE The MediSure Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger It is to be used as an aid in m...

Page 5: ...each patient This Blood Glucose Monitoring System may only be used for testing multiple patients when Standard Precautions and the manufacturer s disinfection procedures are followed Only auto disabl...

Page 6: ...Storage Conditions Temperature 20 50 C Humidity RH 80 Memory Capacity 960 test results with respective dates and times Measurement Unit mmol L Display LCD display Reaction Time 6 Seconds or less Meter...

Page 7: ...side of test strip box or vial System Operating Conditions Test Strips Use only Medi Sure Blood Glucose Test Strips Operating Conditions Temperature 10 40 C Humidity RH 80 Sample Type Capillary whole...

Page 8: ...Power Off Right Arrow Button Temperature Strip Port Insert Test Strip Blue Side up Main Button Power on Memory Recall Left Arrow Button Show Time Alarm Set LCD Display Blood Glucose Test Results Mess...

Page 9: ...7 Meter Back MediSure Blood Glucose Test Strip...

Page 10: ...8 How to determine whether blood sample is sufficient NOTE If the sample is sufficient the sampling window on the test strip should be full with no white space showing...

Page 11: ...he battery symbol appears on the display it means your battery is low Although the meter will still provide accurate results the battery should be replaced as soon as possible After the battery is rep...

Page 12: ...will power on and prompt you to insert a test strip The meter will stay on for 1 minute Or Insert a Test Strip into the meter The Meter will power on and prompt you to take blood sample The meter will...

Page 13: ...hold the button for 3 seconds The year flashes on the screen Press the or buttons until the correct year is displayed Press the button again the month flashes on the LCD screen Press the or buttons un...

Page 14: ...ss the or buttons until the correct day is displayed Press the button The hour flashes Press the or buttons until the correct hour is displayed Press the button The minutes flash Press the or buttons...

Page 15: ...All date and time settings will be saved NOTE To quickly adjust any function keep pressing the or buttons until the one you want appears At any point if you want to shut the meter off press the butto...

Page 16: ...ns 4 Press the button again the hour flashes on the LCD screen 5 Press the or buttons button until the correct hour is displayed 6 Press the button The minute flashes 7 Press the or buttons until the...

Page 17: ...eter 2 When the test strip symbol flashes press the button again 3 Your last test results in memory will be displayed the latest test result will be shown for one minute The number in memory e g 001 w...

Page 18: ...own on the screen for 1 minute 4 Press the button again the 21 day average result will be shown on the screen for 1 minute 5 Press the button again the 28 day average result will be shown on the scree...

Page 19: ...urn off automatically At any point in Memory mode if you want to shut the meter off press the button the meter will turn off and all settings will be automatically saved Results from tests with contro...

Page 20: ...shown on the screen for 3 seconds and the meter will turn off automatically 3 If the alarm has been set the alarm symbol will be shown on the screen 4 Press the button and the temperature and thermom...

Page 21: ...eps if you want to shut the meter off press the button for about 3 seconds and all setting will be automatically saved Alarm Notes 1 The alarm duration is 30 seconds 2 You can press any button and the...

Page 22: ...ES Only an auto disabling single use lancing device should be used A new pair of clean gloves should be worn by the user before testing each patient Please refer to the section of How to Clean Your De...

Page 23: ...ou have 3 minutes to apply the blood sample If you do not apply the sample within 3 minutes the meter will turn off automatically 5 Place the lancing device against the fingertip and press the trigger...

Page 24: ...Log Book 11 Remove the test strip from the meter Discard the used test strip and lancet in a puncture proof container CAUTION When a result either falls in a hypoglycemic range 2 8 mmol L or in the hy...

Page 25: ...tion If you do not apply the control solution within 3 minutes the meter will turn off automatically 5 Use a clean wet tissue to wipe the strip vial cap top Dry thoroughly 6 Roll Control Solution Bott...

Page 26: ...r detailed information please refer to the Glucose Control Solution instructions When Control Solution Tests are Recommended A new Vial of test strips is opened You receive a new lot of test strips Yo...

Page 27: ...spense CaviWipes towelette 2 Clean surface of your device with towelette to remove debris and bio burden 3 Discard used towelette STEP 2 Disinfect After Cleaning 1 Dispense CaviWipes towelette 2 Disin...

Page 28: ...ed and disinfected has been tested for up to 18 250 cycles The recommended frequency to clean and disinfect in a multi patient setting is after each patient 4 If the meter is being operated by a secon...

Page 29: ...tant step in these regulations is to prevent a blood borne pathogens being transmitted if there is shared use of a medical device e g Blood Glucose Meters and test strips The use of a 10 bleach soluti...

Page 30: ...28 Parts of the medical devices that require cleaning and disinfection 1 Whole Case 2 Display Cover 3 Buttons WE RECOMMEND THAT THE ENTIRE DEVICE BE DISINFECTED...

Page 31: ...tis C Virus HCV and Human Immunodeficiency Virus HIV 1 The time to kill above viruses is 2 minutes Diisobutyl phenoxy enthoxyethyl dimethyl benzyl ammonium chloride is the active ingredient for using...

Page 32: ...isinfect Your Device 4 Do not wash or pour liquid on the meter Handling Storage 1 Handle your meter with care and protect it from direct sunlight or extremely high or low temperature 2 Do not expose y...

Page 33: ...r turns on automatically No Memory in the Meter Message If there is no memory in the meter you press button two times and the meter will show 0 Mem Sample Application Message This requires you to appl...

Page 34: ...nging from 1 1 to 41 7 mmol L with respective dates and times HI Message If the test result is higher than 41 7 mmol L the HI message is displayed Lo Message If the test result is lower than 1 1 mmol...

Page 35: ...e was applied before inserting the test strip 4 The blood sample or control solution was applied without the blood drop or control bottle symbol flashed Action 1 Do not remove the test strip before th...

Page 36: ...tion Move your meter and test strips to a place where the temperature is appropriate for running a test Wait at least 30 minutes and test again Cause The battery is low Action Replace the battery soon...

Page 37: ...ED VALUES1 Expected blood glucose levels for people without diabetes Time Range mmol L Fasting 7 0 Two Hours After Meal 11 1 1American Diabetes Association Diabetes Care Volume 35 Supplement 1 January...

Page 38: ...or workmanships Medi Sure Canada Inc will replace It with an equivalent product or repair it at no cost to the purchaser This warranty does not apply to the battery supplied with the meter or to a me...

Page 39: ...37 NOTES...

Reviews: