15
Blue LED flashing
accompanied by
long (1s) beeping
pattern
The
environmental
temperature is
too high for the
battery to
operate safely
Ensure that the battery is placed
where there is no excess heat. Do not
rest the battery on top of radiators or
electronic devices that may emit heat.
Do not attempt to use the battery in
temperatures exceeding 40° C / 104° F
Blue LED flashing
accompanied by
short (500ms) and
long (1s) alternate
beeping pattern
Incorrect AC
adapter being
used
Pilot-12 Lite: use the AC adapter
provided with your Respironics
DreamStation or Respironics System
One S60 or other compatible
power adapters (see section 1.3).
Pilot-24 Lite: use the AC adapter
provided with your ResMed S9 or
ResMed AirSense 10 or the Medistrom
AC adapter (see section 2.3).
PAP device
connected
to battery is
incompatible
Ensure that the PAP device you are
using is compatible with the Pilot-12
Lite or Pilot-24 Lite (see sections 1.3
and 2.3 respectively)
Four green LEDs
flashing together
accompanied by
long (1s) beeping
pattern
Incorrect AC
adapter being
used
Pilot-12 Lite: use the AC adapter provided
with your Respironics DreamStation or
Respironics System One S60 or other
compatible power adapters (see section
1.3)
Pilot-24 Lite: use the AC adapter provided
with your ResMed S9 or ResMed AirSense
10 or the Medistrom AC adapter (see
section 2.3)If problem persists please
contact Medistrom technical support at
1-888-407-4791 (North America), 1-300-
146-358 (Australia) or email support@
medistrom.com
Blue LED flashing
accompanied by
a short (500ms)
short (500ms)
long (1s) beeping
pattern
PAP device
connected
to battery is
incompatible
Ensure that the PAP device you are
using is compatible with the Pilot-12
Lite or Pilot-24 Lite (see sections 1.3
and 2.3 respectively)
DC output cable
is damaged
Ensure that the cable connecting the
Pilot-12 Lite or Pilot-24 Lite to your
compatible PAP device is not damaged
Blue LED flashing
accompanied
by three short
(500ms) beeping
pattern
PAP device
connected
to battery is
incompatible
Ensure that the PAP device you are
using is compatible with the Pilot-12
Lite or Pilot-24 Lite (see sections 1.3
and 2.3 respectively)
Blue LED
periodically
flashing (no
pattern)
Maximum battery
power output
No action required, the battery is
operating correctly.
Depending on
your prescribed pressure, enabled
PAP device features and breathing
rate the battery may reach maximum
power output which is indicated by the
periodically flashing blue LED
Blue LED flashing
every 0.5 seconds
Battery capacity is
below 5%
Charge the battery as soon as
possible.
The battery power level is
critically low. Not charging the battery
for a prolonged period will cause it to
go into Over Discharge State.