Fault Cause
Remedy
LED fails to light up
No battery in the device Replace battery
or battery spent
Battery not inserted
Insert battery again
correctly
Device not switched on
Switch device on
Insulating film has not
Remove insulating film
been removed
from between battery
and device
LED lights up weakly
Battery spent
Replace battery
Low output from device
Electrode pad dirty
Clean /replace electrode
pads
Device not applied
Re-apply device
properly
LED function normal
Battery spent
Replace battery
Electrode pad dirty or
Check electrode pads
short-circuited
and replace if necessary
Device switches off
Battery spent
Replace battery
Electrode pad worn
Check electrode pads
out or dirty
and replace if necessary
Device not applied
Re-apply device
properly
Electrodes:
• Clean the adhesive surface after use to achieve better adhesion. Moisten the gel
with a drop of water and allow it to dry briefly. Then stick the electrode back on
the protective film.
• The length of time the electrodes remain usable depends on their storage, skin
type and use. Store the electrodes in their storage box in a cool, dry place.
Device:
• The device does not require any special maintenance.
• Remove the battery before cleaning the device.
• Never use aggressive cleaning products or stiff brushes.
• Clean the device with a soft, possibly slightly moistened cloth. Do not
allow water to penetrate the device. Do not use the device again until it is
completely dry.
• Remove the battery from the device if it is not going to be used for some
time.
• Do not expose the device to direct sunlight and protect it against dirt and
moisture.
• With the electrode pad on the protective plastic film, store the device in a
cool, dry place in the storage box provided on purchase.
5.2
Cleaning and
maintenance
5.1
Trouble-
shooting
but no power output
prematurely
Please contact your
service centre if it
is not possible to
remedy a fault in the
manner described.