Operation
SI PCF User Manual
Ref: DOC03148 Issue 1 Friday, 06 January 2017
Page 39 of 81
Power Up
Turn on the power to the machine using the
isolation switch on the wall.
When the machine is turned on, it conducts a
Power-On Self-Test (POST) to ensure the
machine is ready for use.
This process takes a few seconds to configure
the machine and check the basic operation of
the control system, memory and sensor
interfaces.
Authorisation code
This is an optional feature, which can be
enabled by a service engineer. When enabled,
the user is required to enter a four-digit PIN
number to unlock the user control panel. A
valid PIN entry identifies the user and their
operator number is shown on the cycle
printout. There are 80 operator numbers each
with a unique PIN number, see Appendix 6-8
for details.
Setting Time & Date
The time and date must be set correctly to ensure the accuracy of the cycle records, and
MUST
be
periodically checked by the user. There is no automatic provision for daylight saving hours. While the
machine is off, a backup battery maintains the time and date.
Set Time
1. Use the scroll arrows to cycle
through to the clock icon
2. Press
Set Time
3. Use the arrows to adjust the
Hours, Minutes and Seconds
4. Press
to return to the main menu
Set Date
1. Use the scroll arrows to cycle
through to the clock icon
2. Press
Set Date
3. Use the arrows to adjust the
Date, Month and Year
4.
Press
to return to the main menu