28
29
CONNECTION WITH CHARGING CONNECTORS WHICH ARE
A PART OF THE CHASSIS/MONOBLOCK
CHARGING PARAMETERS AFTER CONNECTION TO THE AMBULANCE ELECTRICAL
SYSTEM
Voltage to activate the system and start
charging – with the chassis/monoblock
system switched off
13.7 V
Voltage range to start charging – with
chassis/monoblock system switched on;
Limit voltage to stop charging
9.0 V – 15.0
V *
Maximum charging current:
4 A
* Adjustable using the service application
WARNING
The limit voltage value to start charging can be set only by a trained person or
an authorized partner. There is a risk of damage to the product.
CONTACT
GND (-)
CONTACT
12V (+)
Charging in the Ambulance
The rear fixing mechanism includes charging contacts. Charging starts automatically
when the chassis/monoblock is inserted into the fixing mechanism and connected to
the charging connectors. After the contact is established, charging will start within 10
seconds.
For more information about the battery charging function, refer to the Chassis/
Monoblock User's Manual.