WARNING
Always make sure that the rear horizontal and front vertical pins are completely
locked. The fixing lever must be fully raised to the horizontal position.
Improper locking of the fastening system can lead to serious injuries.
FIXING LEVER IN THE WRONG POSITION
REAR PIN OF THE CHASSIS/
MONOBLOCK
FRONT PIN OF THE CHASSIS/
MONOBLOCK
FIXING LEVER IN THE LOCKED
(HORIZONTAL) POSITION
HORIZONTAL PIN
IN LOCKED POSITION
VERTICAL PIN
IN LOCKED POSITION
Exiting the Ambulance
Step 1
To unlock the fixing mechanism, press the fixing lever. The horizontal pin and the pair
of vertical pins retract and lock in the unlocked position. Slide the chassis/monoblock
out of the ambulance.
After moving to the limit position, the auxiliary spring is pushed, which prevents the
unintended return movement towards the ambulance. When you push slightly towards
the ambulance, you will feel a slight resistance of the spring, which will make you sure
that it is in the limit position. Before the next step, always make sure that the chassis/
monoblock is completely in the limit position.
2.
1.
FIXING LEVER
WARNING
After the loading device is pulled out of the ambulance, the auxiliary spring acts,
which prevents backward movement. However, this is not sufficient protection
against movement due to external forces in real conditions. Never leave the patient
unattended.
NOTE
After pulling out the chassis/monoblock or rather disconnecting the charging
connectors, the undercarriage is automatically pulled and at the same time the control
lever of the sliding arm is pressed. This ensures a smooth exit from the ambulance.