Record all the control solution test results with CONTROL
SOLUTION ICON ‘ ’ to distinguish them from those of
actual blood glucose tests in the meter memory. Recorded
control solution test results will not be calculated in your
result averages.
To record the control solution test results, press any of +/- button of the meter
to select CONTROL SOLUTION ICON. Press and release the power button if
selection is completed. The control solution test results will be saved.
Record the Control Solution Test Results in the Meter
If your control solution test results continue to fall outside the range printed on
the test strip vial, the GlucoDr auto Blood Glucose Monitoring System may not
be working properly.
- If this is the case, do not use the system to test your blood glucose levels.
- Contact your local distributor.
1